EONE ET201, a 2-in-1 multimeter, or how they crossed a snake with a hedgehog
The packaging is extremely plain. The only information on it is the instrument model and a barcode sticker.

Inside the package are a carrying case with the instrument and a fairly substantial manual. Because of its size, the manual isn’t in the case—which would have made more sense—but is packed separately.

The case contains the multimeter itself and everything that comes with it.

I wouldn’t call the package contents extensive:
1. Multimeter-oscilloscope
2. Test lead set
3. Adapter for testing transistors
4. Carrying case
5. Manual
At the very least, alligator clips would have been useful.

The case deserves a separate mention. In my opinion, it’s a bit too good for this class of instrument. I know that sounds a little odd, but despite the multimeter’s relative simplicity, they’ve included a very handy case. For example, I have a UniT 61E, which didn’t come with a case, though one would have been very useful. There’s a compartment for the instrument and test leads, plus a separate pocket. I think that’s excellent. The strap is a bit flimsy, though.
The manual is in English, but it covers literally everything, and in quite a lot of detail. If it weren’t in English, I’d give it five stars; as it is, unfortunately, four at most.

Here’s a chart of the specifications. Most of it is clear without translation, but I’ve loosely “translated” the main specs :)

The test leads are very budget-grade. They’re soft, which is probably the only thing I liked about them. They’re marked CATIII 1000 V, but the leads themselves are long, and the part that plugs into the instrument has no additional insulation.
They also include an adapter for testing transistors. It’s useful in itself, but extremely awkwardly designed: it’s only convenient for testing transistors in TO-92 packages, and even then only with a particular pinout. The popular C945 transistor, for example, comes in two pinout variants, CBE and BCE. Testing the second variant will be very awkward, not to mention transistors in TO-220 packages—their leads simply won’t fit into the holes.

And here’s the little instrument. Though judging by its dimensions, there’s nothing “little” about it :) It’s big, thick, and fairly heavy. Just picking it up, my first impression was: this is a solid piece of kit. The case plastic feels quite nice, and it’s comfortable to hold. But I found some drawbacks while using it, which I’ll get to later.

The upper part of the case has a fairly large 128×64-pixel matrix display, with two rows of buttons below it:
DIS - Switches between multimeter and oscilloscope modes
Up/down arrows - Select memory locations.
SELE - Auto-calibration in oscilloscope mode. In memory mode: press and hold to select a location to clear; short press to delete the data.
POW - Power. Short press to turn on; when on, short press toggles the backlight, and a long press turns the instrument off.
DC/AC - In current or voltage measurement mode, switches between DC and AC.
REL - Relative measurements.
H - Memory functions. Short press saves the reading to memory and displays it at the same time; long press switches to viewing the saved data.
As for the memory functions, they’re awkward to use, and I don’t see much point in them anyway. Compared with my Uni-T 181, where all these functions are genuinely useful and get used—especially the data logger mode—this implementation feels rather clumsy.

Mode selector.
Yes, there’s no autoranging, and in my opinion that’s actually a good thing. A regular selector is much more convenient than a bad autoranging system, while a good one is unlikely at this price and with this feature set.
Besides, after many years I’m used to selecting the range manually, though that’s just a matter of habit. It’s also a little easier to use because there are no separate ranges for AC current and voltage; you select AC with a separate button, which is genuinely convenient.
Overall, I like it, but there’s one downside: the instrument beeps whenever you switch modes. It’s incredibly annoying—I can’t even describe it. That’s how you ruin a good product with an unnecessary feature. And for obvious reasons, disabling the beeper isn’t an option.

Terminals for connecting the test leads.
I don’t know what the designers were smoking, but they really went out of their way to arrange the COM and measurement terminals like this. There’s a sort of “standard” for terminal order around the world: A, mA, COM, V. But the designers here showed some “ingenuity” and swapped the two right-hand terminals. I’ve been caught out by this once before: when measuring current, I naturally plugged the leads into the first and third jacks, then spent ages wondering why there was no current in the circuit....

The meter is quite thick; for comparison, here it is next to a matchbox. I wouldn’t say that’s good or bad, but it does look rather chunky. That said, I have no complaints about the case or the overall build quality—it’s all quite neat.

One unusual feature for a multimeter is the IR remote control tester. It actually detects the 38 kHz carrier used to modulate the IR remote’s commands. It’s quite useful, and more convenient than checking with a phone. If the manufacturer had also added a way to display the transmission protocol—for example, showing RC5—it would be even more useful. And don’t even get me started on displaying the command code. Ah, dreams. I used to use a gadget called URCR, but my computer doesn’t have a COM port, and my USB version can’t work this conveniently.

The stand attaches with small magnets. It takes some getting used to, but I find it quite practical.

It is a little flimsy, though, and lacks rigidity, so it can flex. I also have to mention that the meter has no feet. On a smooth table it slides better than figure skaters on ice. Why they couldn’t stick on a few pieces of rubber is a mystery shrouded in darkness.

The meter runs on four AA cells and works perfectly with four rechargeable batteries.
This is much more convenient than a 9 V battery, but if they’d used three cells, you could replace them with a lithium battery. On the other hand, batteries are usually sold in even-numbered packs, so you can buy a pack and charge them all in the meter. Overall, I think the advantages outweigh the disadvantages.
I should also mention how the cells fit: I’ve never encountered springs this stiff before. You practically have to hammer the cells into place, and the cover is a tight fit too. At least they used proper screws with nuts in the plastic instead of self-tapping screws.
The meter draws around 20–25 mA, depending on the operating mode and whether the backlight is on. Typical 2000 mAh rechargeable batteries will last a long time, especially since both the backlight and the meter itself switch off automatically.

When I first picked up the meter, I also noticed that it had a fair bit of heft.
For comparison, here are photos of the meter under review and some others I weighed earlier.
All the multimeters in the photos have their batteries installed; the one under review has rechargeable batteries. At just 130 g, it came in below my Uni-T 181, which has a built-in 2000 mAh, 7.4 V battery. So it’s definitely not pocket-sized :)

Before testing, I did the obligatory teardown to
The meter is very easy to take apart: four self-tapping screws, and that’s it. The cover is connected to the main body by four wires—two to the battery and two to the buzzer. The wires are soldered in place.

What can I say? It looks quite neat, and I like it overall.

Most of the components are concentrated in the upper part of the case. The soldering is acceptable, though residue from poor cleaning in a few places keeps it from being really good.

The range selector board is a separate piece, and it’s clean and neatly made. The small screw in the center adjusts the force needed to turn the knob.

I have no major complaints about the terminals either, except that they look a little undersized for the claimed 20 A. Otherwise, they’re just ordinary terminals—I’ve seen both better and worse.

The selector board connects through several connectors and is held in place by three self-tapping screws. Even after removing them, the board is very difficult to pull out.

Unusually, the meter has three fuses instead of two. I assume that’s a precaution because the jacks are laid out incorrectly. If only they’d included spare fuses...
I’m glad to see a proper fuse in the current-measurement circuit, rated for up to 20 A. That’s genuinely good. I wouldn’t recommend blowing it, though—these fuses are relatively expensive.
The two small ones are identical, rated at 400 mA.

Judging by the number of contacts on the selector, I’ll venture that the switching is electronic and the selector simply chooses the required mode. But this is just a guess.
The switch is implemented properly; the moving contacts don't slide across any "unnecessary" traces.

It also has protection, which is a nice touch.
1. 1100 V varistor
2. Several TVS diodes.
3. There are also three thermistors on the board. You can see a crack near the lead on the left; that's most likely my fault—I may have damaged it while removing the board.
4. This, however, is "not my doing"; you can see signs of poor soldering and flux.

They didn't come up with anything else, so improvise (C) Radio Day.
Just a photo.

Let's take the board out of the case. I strongly recommend against doing this; you'll be fed up with blowing dust out of the window afterward.

The top eight buttons are actual buttons, which is a nice touch. Right at the beginning, I noticed the unusual clicks when switching. In my opinion, it's a quality design choice—if only they'd cleaned off the flux.
The meter has a dual shunt for the 20 A range. This may be a good solution, since the current is quite high and using two shunts should make operation more stable by keeping their temperature lower.

1. The multimeter's main chip is a glob-top, but it's mounted on a small board that's soldered to the main board. In my opinion, this is a little more reliable than putting the glob-top directly on the main board, since it's less susceptible to thermal deformation.
2. Some kind of op-amp next to it.
3. There's also an 8-bit UPD78F0537 microcontroller from NEC on the board. I assume it handles the matrix display and oscilloscope functions.
4. To the right of the microcontroller is an EEPROM chip. Its purpose is unknown; presumably it stores calibration data.

There are also several logic chips and a couple more op-amps scattered around the board.

An IR receiver is located at the front of the case. It's soldered in place well above the window, but this doesn't affect its operation at all.
The switch below it caught my eye. I hoped it would turn off the switching sound, but it turns out it puts the meter into calibration mode. For obvious reasons, once I saw CAL on the screen, I didn't do anything else.

Display.
As I mentioned above, this meter has a 128×64-pixel matrix display. That's more than enough for this application. A TFT like the one in my Uni-T 181 would certainly be better, but it would need a more powerful processor to handle the data, which would push up the price of the meter.
The resolution is great, of course, but visibility is another matter... I don't know, I personally found it uncomfortable. Let me explain.
1. Under normal overhead lighting, the readings are visible but hard to read.
2. I turned on a desk lamp to the side. That helped, but because the display is recessed so deeply, a shadow appeared on the left.
3. With the backlight on, it's very easy to see, but although it uses little power, it has one drawback: it turns off after 30 seconds, and having to turn it back on is annoying.
4. It's quite easy to see with the flash on, but it's better to press the button every 30 seconds.
So, in bright sunlight or good lighting, there's no problem; everything will be fine. But in poor lighting, there are only two options: keep turning on the backlight every half minute or squint at the readings. Not convenient.

That's it. Now for a few tests. Since I don't have a calibrator or reference components, I'll compare its readings with those from a meter with a known higher accuracy class.
But first, a quick check of a few parameters unrelated to measurement accuracy.
1. The meter's input resistance is about 11 MΩ.
2. Voltage across the meter's terminals when measuring resistance: 0.61 V
3. Voltage across the terminals in continuity mode: 3.08 V
4. The short-circuit current is 1 mA
There are usually six items here, but since the meter combines the semiconductor test and audible continuity functions, there are only four.
The audible continuity test is fairly quick, although there's a slight delay.

Next, a test against the voltage reference. Since the meter can display a maximum of only 1999, I tested it on the 20 V range.
The readings are all slightly low, by almost the same number of counts in nearly every case. Overall, though, they're quite respectable.
By the way, the photo shows that sometimes the top and bottom halves of the screen have different contrast. This is a result of the "scanning" process: the processor refreshes the entire screen. You don't notice it during normal use; it only shows up in photos.

Comparative DC voltage measurement. The first three checks are at the top of the range, and the fourth is almost at the maximum output of my power supply.
The 200 mV and 2 V ranges are spot-on. The 20 V and 200 V ranges read a little low, but overall they’re still good.
By the way, as you approach the limit of the selected range, the meter starts beeping. At relatively high voltages, it gives a short beep and displays a lightning bolt icon.

Let’s see how the meter does measuring current. It was during this test that I was trying to figure out why nothing worked—until I realized I’d connected the probes incorrectly.
As you can see, there are no issues here either.
For the AC voltage source, I used a regular mains outlet. The readings also matched my multimeter (last photo).

Capacitance measurement.
Up to 100 nF, there are no issues: the readings are accurate. Above that, it starts reading a little high.
Unfortunately, though, the upper limit is just 200 µF, which is a bit low. Even inexpensive models can measure more these days.

Resistance measurement.
The only minor issue is on the 200 Ω range, but I’d put that down to the rather poor quality of the supplied probes. Also, I took these measurements without pressing the REL button.
Otherwise, the error is within the last digit, which is pretty good in my opinion.

Now we come to this multimeter’s special feature: its oscilloscope function. To be fair, though, calling this mode an “oscilloscope” is a bit of a stretch. It’s more of an oscilloscope probe, with a frequency limit of 10 kHz.
Mostly for the sake of experimenting—and to demonstrate the feature—I connected the multimeter to the output of my oscilloscope, which was running in generator mode.
Incidentally, you also have to select the voltage range here, just as you do for regular measurements. So, to measure 5 V signals, select the 20 V range, and so on.
Let’s feed it a square wave at 10, 100, 1,000, 5,000, 10,000, and 20,000 Hz.
The waveform appears on the screen. If you get something incomprehensible instead, there are three possible reasons:
1. The wrong range is selected
2. The frequency is above 20 kHz
3. You need to press the SELE button.
The signal’s voltage and frequency are shown to the right of the waveform. Frequency measurement is quite coarse, with an upper limit of 19.99 kHz, so at 20 kHz it displays oL for overload.
The square wave looks reasonably clear at 10 kHz. At 20 kHz it’s a mess, and I can’t select any frequencies in between.

The triangle and sawtooth waves look good at 1 kHz. At 10 kHz, you can just about make out the waveform.

Sine waves at 100, 1,000, 5,000, and 10,000 Hz. There are no problems up to 5 kHz. At 10 kHz, the display is noticeably worse, but still within the stated 10 kHz limit.

Next, a few tests outside normal operating conditions.
1, 2. I tried viewing waveforms at 200 kHz and 1 MHz. As you’d expect, all we get is a mess: the frequency counter shows nonsense, and above 1 MHz there’s no waveform at all.
3, 4. Next, I switched the multimeter to frequency measurement mode—Hz on the dial. But the meter clearly has no idea what I want from it. More accurately, the 1.5–2 V I fed it just isn’t enough, so it keeps displaying whatever it feels like. The actual input signals are 5 V and 1 kHz, respectively.

But when I plugged it into a wall outlet, everything worked as it should: 50 Hz. It’s odd that a meter which measures frequency quite well in oscilloscope mode won’t measure it in the dedicated mode. The most likely reason is that the meter is designed to handle a wide range of input voltages, which results in poor performance at low voltages.

A few more features.
As I mentioned above, the multimeter can measure the gain of bipolar transistors. But this function also has some limitations. To start with, the adapter itself isn’t very convenient, as I mentioned earlier. And the maximum range is 2000, so forget about testing Darlington transistors—you’re limited to ordinary bipolar transistors.

And finally, the service functions.
The meter has some memory for storing readings: 3 slots for waveforms and 100 for other values.
To save a reading, press the H button while measuring. For example, H2 will appear, where 2 is the memory slot number.
Press and hold H to view the saved readings. If you press SELE here, a trash can icon appears (it took me a while to figure out what it was); press it again to clear the slot. For some reason, though, it cleared all the cells.

I put all the test data into a small table. It’s a little rough around the edges, of course, but it gives a very clear overall picture: the meter passed most of the tests with plenty of room to spare.
The thing is, almost everywhere the specification says ±10 counts, and it only exceeded the percentage tolerance right at the end of the capacitance measurement.

Now I’ll try to sum things up.
So, what are we looking at? A hypertrophied version of the very popular Mastech M890 multimeter. In other words, the manufacturer took a basic meter, added a matrix display, an oscilloscope function and memory, and came up with the device reviewed here.
Most of the complaints are really about the design and display quality. I wouldn’t criticize it for measurement accuracy, since it measures quite well. But meeting such modest specifications isn’t particularly difficult.
The idea itself isn’t bad. Sometimes it really is handy to have a “jack of all trades,” but in my opinion the manufacturer was a little late with this approach—or, more precisely, with these specifications. If this had come out 10 years ago, I’d have said it was very good; 5 years ago, not bad, but lacking a bit. Today, though, the specs are rather modest.
That’s because meters with 4000 counts have since appeared on the market, and they also offer wider capacitance and frequency measurement ranges. And let’s not forget the DIY oscilloscopes, where a modest $18–20 gets you a TFT display and a 200 kHz input signal frequency, rather than 10 kHz as here.
The power supply is implemented quite conveniently: four batteries or rechargeable cells have much more capacity than the familiar “9V batteries.” With auto power-off, I think it should last a long time.
The included pouch is handy too. It’s a small thing, but a nice touch. It’s a shame that better-known manufacturers so often “forget” about it.
If the manufacturer continues developing this product, perhaps the next step will be
to “crossbreed” newer multimeter chips with that same 200 kHz oscilloscope.
And of course, there are plenty of things to criticize. There are no feet, so the meter slides around. The screen is hard to read; a way to keep the backlight on continuously would have made it fine. The beeper when switching ranges is annoying. Working with the memory is awkward and annoying, though not as much as the beeper.
Where can you use this meter? In my opinion, either on a hobbyist’s workbench or in specialized applications where versatility matters and the oscilloscope’s 10 kHz measurement range is enough. Such applications certainly exist, but I don’t know much about them, since I work with power supplies where 10 kHz is nowhere near enough. I’d be more likely to buy a basic multimeter for $15–20 with better specs, plus a simple oscilloscope for another $20. The total would be about the same, but the measurements would be more accurate. It would just be less convenient to use.
In short, it’s a solid, fairly well-built device with some shortcomings and good accuracy, but it’s clearly “behind the times” in terms of specifications, even though the idea itself is interesting.
Video version of the review—or, more accurately, the inspection.
https://www.kirich.blog/obzory/izmeritelnoe/350-eone-et201-multimetr-2-v-1-ili-kak-skrestili-uzha-i-ezha, et120mc2 инструкция, et120mc2 инструкция на русском, инструкция для осцилографа-мультиметра et 828 pro на русском языке, мультиметр -осцилограф et826 инструкция на русском, тестер ет 826 инструкция на русском, мультиметр ет-201, et828 тестер отзывы пользователей, et 826 осциллограф отзывы, руководство пользователя мультиметра et8134 на русском языке, ет 826 2в1.200к инструкция по применению, ет 828 мультиметр инструкция на русском, manual осциллографа et 310a, руководство пользователя et 826 на русском, et826 мультиметр инструкция, мт8206 мануал, woopower et828, инструкция et8134 мультиметр на русском, руководство на русском языке осциллографа ет120м, мультиметр et828 мануал на русском, et 826 инструкция, отзывы mustool mt8206, осциллограф ет826 инструкция на русском языке, et828 отзывы, инструкция на русском для тестера et828
Questions and answers
What can the oscilloscope in the EONE ET201 do?
The built-in oscilloscope handles input signals up to 10 kHz. The author says this is far too little for working with power supplies.
What are the EONE ET201’s drawbacks?
It slides on a smooth table, the display is hard to read, and the beep when switching modes is irritating. The author also finds the memory controls awkward.
What batteries does the EONE ET201 use?
It runs on four AA cells and, according to the author, also works well with four rechargeable batteries of that size.
This model is included in the summary table: Comparison of measuring instruments reviewed on the blog
The figures from this review are in the measurement database: Measuring instruments