Owon CM2100 clamp multimeter with standalone data logger
A reminder: to keep up with new posts on the blog, I recommend following me on Instagram. I’ll post notifications about new articles there and might also share something interesting from time to time - link to my account
You can also subscribe to updates and new articles on the Telegram channel - https://t.me/KirichBlog
First, it’s worth mentioning that Owon is a fairly well-known company, and I’ve already reviewed one of their multimeters. Of course, it’s no Fluke, but it’s definitely not some no-name brand either, so their products are worth a look.
There’s a sale on right now, so the basic version costs $32.42, while the Bluetooth version is $38.95. At that price, it’s worth buying.
Anyway, I ordered a pair of these clamps. They were packed neatly and come with a case. But why “Blluetooth” is spelled with an extra L remains a mystery to me...
There are two versions, the CM2100 and CM2100B. The latter has a Bluetooth module and costs more. If you don’t need it, the regular version is the better buy, although I’d recommend considering the Bluetooth model in particular.

The package includes:
1. CM2100/CM2100B clamp meter
2. Test leads
3. Screwdriver
4. Manual
5. A sheet with a link to the software
6. Case.
Batteries are not included!

The manual is in English, but it’s very detailed and has lots of pictures, so even a basic knowledge of technical English should be enough to understand how to use the meter.
Interestingly, one of the lines on the specifications page has been covered up. It clearly shows a voltage measurement range up to 20 mV. I wonder if there’s a model that can measure that?

A simple, budget screwdriver, but useful.
The test leads are nothing special either. They have protective caps, which give them a CAT-III 600 V rating; without the caps, they’re rated CAT-II 1000 V. The cable cross-section isn’t specified, and there’s no other information printed on them. The only rating given on the leads themselves is 10 A.

The device is compact, with a rubberized, non-slip casing. There’s a protective film on the front panel that’s easy to miss. It’s blue, like most instruments from this company.

On top are the rotary mode selector, three buttons, the display, and an LED to the left of the display for NCV mode (non-contact voltage detection). The LED isn’t labeled at all, so at first I couldn’t figure out what it was for.

The LED is marked in the picture, but who looks at the pictures before using the device? :)

The clamp jaws are at the top. They’re marked with the electrical safety ratings CAT-II 600 V and CAT-III 300 V, as well as the polarity for positioning a wire when measuring DC current. Even so, I still couldn’t figure out how to tell which way the wire should pass through.
There are two test lead sockets on the opposite end of the casing, with the same ratings: CAT-II 600 V and CAT-III 300 V.

The battery compartment is on the bottom. The meter runs on two AAA batteries. It’s thoughtful of them to include a screwdriver for opening the cover.

When you turn the meter on, all the segments light up briefly. I can say that almost all of them are actually used. The meter then switches to normal operation. The readings are very easy to read: the characters are sharp and high-contrast. The horizontal viewing angle is almost all the way around the casing; the vertical angle is noticeably worse, but I wouldn’t call it a serious issue.
The display has 2,000/20,000 counts, depending on the measurement mode: 20,000 in voltage mode, for example, and 2,000 in current mode.

Speaking of the display, it’s oriented along the length of the casing. I also use a pair of clamp meters that are complete clones of the Mastech MS2108, in a more “classic” design. Their casing is noticeably larger, yet the display characters are smaller.

1. The NCV sensor is built into one of the jaws and looks like a raised bump.
2–4. The display isn’t particularly large, but the meter only measures up to 100 A, so it’s actually more than adequate for the current range.

The mode selector is a little stiff, though that’s just my subjective opinion.
Voltage measurement
Current measurement, with three ranges: 2, 20 and 100 A.
NCV function
Resistance and capacitance measurement, semiconductor testing and continuity testing. The capacitance and resistance symbols on the selector are swapped, though: when you select this mode, it starts with resistance measurement, not capacitance.
Frequency/duty cycle measurement
The two end positions are both for switching the meter off, which is convenient. It also has an auto-off function that kicks in after 15 minutes.
There are three buttons next to it:
Select/V.F.C — press briefly to select a mode; press and hold to enable the V.F.C function
Zero/BT — zeroing/Bluetooth on
Hold/Backlight

The description includes a list of all supported functions.

The backlight is even and not too bright, but makes the readings easy to see. The display is also perfectly readable in dim light without the backlight.

Now, a little about the modes.
1, 2. DC and AC voltage measurement; the range is selected automatically.
3. In every mode, you can freeze the readings on the display.
4, 5. The VFC function is enabled in a rather odd way: it displays the start of operation, then you take a measurement, and when you turn it off, it displays the end of operation.
6. Bluetooth is enabled by pressing and holding the blue button. By the way, enabling this mode disables APO (auto power-off), which is necessary for long measurements/logging.
7–8. DC/AC current measurement.
9. Zero function, which manually zeros the readings. It works in several modes (voltage, current, resistance and capacitance), but is most useful for DC current measurements. In the second-to-last photo, you can see that without zeroing, the clamp meter reads 516 mA (the current may be different); that’s what you need to zero before measuring.
I still don’t quite understand the VFC function. The manufacturer says it:
+VFC mode can filter high-frequency interference signals to ensure accurate measurement data
One of our local stores describes it this way:
Режим +VFC може фільтрувати високочастотні перешкоди для забезпечення точних даних вимірювань.
That sounds clear enough, but usually this mode simply turns on like a regular filter. Here, though, we see the start and end of operation, much like with the Max/Min function. In any case, I couldn’t find a source where I could observe this function at work. In theory, it should improve the accuracy of readings, for example when measuring current in brushed motors.
Yes, this function only works in AC mode.

1, 2. Current measurement on the 20 and 100 A ranges.
3. NCV function: bring the clamp’s “nose” close to a live wire, and the LED lights up and the buzzer sounds.
4–7. Resistance and capacitance measurement, diode testing and audible continuity testing.
8, 9. Frequency and duty cycle measurement.

Tests.
First, some measurements that aren’t really about metrology, but may still be useful.
1, 4. Voltage and current during resistance measurement—the lower, the better.
2, 5. Voltage and current during diode testing—the higher the voltage, the better.
3, 6. Voltage and current during audible continuity testing.
Overall, the results are pretty typical. For example, the well-known Uni-T 61E measured 0.2/1.8, 2.9/1.5 and 3.05/1.5, respectively.
What I liked:
The voltage in diode-test mode: 3.25 V is good, especially considering that the meter is powered by 3 V.
The polarity doesn’t change in audible continuity mode. I don’t know about you, but it annoys me that the polarity on my Uni-T changes between modes...
Speaking of audible continuity testing, I’d say it’s average. It has a slight delay before it kicks in—more precisely, the short-circuit duration has to exceed a certain threshold, around 0.1–0.2 s, for example. The minimum beep duration is also fixed. In use, it reminded me somewhat of the continuity tester on my UT181A.

And now, a little metrology.
First, let’s see what the manufacturer claims. By the way, note that enabling VFC mode increases the maximum error.
By the way, this table shows a 20 mV range, but the actual minimum range here is 200 mV.

Let's start with DC voltage measurement. To check it, I'll use a voltage reference. Just in case, I'll include the results printed on its label (first column) and the results I got when it was measured by a metrology lab (second column).

This measurement usually shows how accurately the multimeter's built-in voltage reference is calibrated, and the meter surprised me a little here. At first I photographed the clamp meter's readings, but since they differed noticeably from the expected values, I decided to double-check with my own multimeter in case my voltage reference had gone bad... But no, the reference is fine; the clamp meter simply reads low.
No, the measured error is well within the stated limits—by a mile, in fact: it's 0.1–0.2%, against a stated 0.5% + 5 counts. Still, it's a bit strange...

Next, measurements using fairly accurate resistors and capacitors. I used resistors with 0.25–0.5% tolerance and capacitors with 1% tolerance, which is perfectly acceptable given the meter's stated error of 0.5–0.8% and 3%, respectively. And bear in mind that these are wirewound resistors, military surplus ones, in fact—I took apart some tank control units a long time ago.
First, 75 and 510 Ω, 1.6, 2, 47 and 180 kΩ resistors.

And 1000, 1800 and 7850 pF, plus 0.1, 0.39025 and 100 µF capacitors.

No issues here; the readings almost match my UT-181A, which I have no complaints about.
And here I come back to the question of the meter's built-in voltage reference. The voltage readings were a little low. If the reference were the problem, the resistance readings would also be proportionally low, but that's not what we see.

DC current measurement is also mostly trouble-free. The only exception is that at the very end I tested at 20 and 24 A without any additional verification. The current was set by the power supply, so those results are questionable. All the other modes look very good.
Almost all measurements followed the 1-2-5, 10-20-50, etc. sequence to cover as much of each range as possible. By the way, I measured at 1 and 2 A on the 20 A range, even though the meter can measure those currents on the 2 A range.

But it turned out not everything was quite so rosy. Just out of curiosity, I reversed the polarity (zeroing the meter again, of course) and checked once more.
No, the results didn't suddenly become bad—not at all—but at very low currents they differ by more than they did the first time. In general, measuring such small currents with a clamp meter is a bit like witchcraft: switch it on, position the clamp exactly as it will be during the measurement, zero it, then take the reading.

By the way, the stated upper limit for current measurement is 2000 counts, but in practice you can measure almost up to 2200; above that, it displays an overload.

To be honest, AC measurements don't concern me much, and here's why: first, it's much harder to mess up a current measurement than an AC measurement, and for AC we measure through the same divider, so the results will essentially be the same.
But I couldn't skip checking the accuracy. I was interested in the stated TrueRMS mode and the claimed maximum frequency of 1 kHz for AC current/voltage.
First, we feed in a sine wave. I applied a signal with an amplitude of 3 V peak-to-peak, so the display shows an RMS voltage just over 1 V.
The test was performed at 50 Hz, 1, 5 and 10 kHz.
The results are correct at 50 Hz and 1 kHz, clearly much too low at 5 kHz, and nothing is displayed at 10 kHz. Then again, nobody promised performance above 1 kHz.

But it's much more interesting to see how the meter handles nonsinusoidal signals. I'll test it at 50 Hz and 1 kHz using square, sawtooth and triangular waveforms.
1, 4. Square wave at 50 Hz and 1 kHz
2, 5. Sawtooth wave at 50 Hz and 1 kHz
3, 6. Triangle wave at 50 Hz and 1 kHz.
Overall, it's fine. The readings are a little low at 1 kHz with square and sawtooth waves, but to be fair, my UT-181A does that too, albeit to a lesser extent. In any case, I can say that it has TrueRMS.

Out of curiosity, I checked a triangular signal at 10 and 20 kHz.

I can already hear the question: since I'm comparing it with the UT-181A, how does its frequency range hold up when measuring AC voltage?
Let's take a look.
1, 2. No problem up to 600 kHz.
3. Above 700 kHz, the readings start drifting upward.
4. At 1 MHz, it starts clearly underreporting the results.
In my opinion, that's quite enough for it to serve as a reference when checking other instruments. Plus, it has a separate hardware chip for TrueRMS measurements.

Frequency measurements are a little low across all ranges, but still within the specified error margin.
There is one caveat: the specs also list an error margin for the 20 MHz range, but the meter won't measure above 3 MHz; at 4 MHz, it displays nonsense.

I checked the signal's duty cycle using a square wave. There's a small error of one least significant digit, but otherwise no issues.

That pretty much covers the measurements. Let's open it up and see how it's put together.

And it's actually assembled quite neatly.
The current-measurement analog section is on the left. You can see three trimmer resistors; I assume there's one for each range.

1. A pair of op-amps SGM8956
2. The main chip's markings have been erased, which is very strange, especially for a relatively large company...
3. Bluetooth module
4. The probes are connected through a PTC thermistor, so that's basically the extent of the protection...

We take the board out of the case. The display stays in the case, but its backlight module comes out.

One of the chips on the underside is a 24C08 EEPROM; the other is marked:
385-12
TI2BM
A495
I couldn't figure out what this chip is off the top of my head. If anyone knows, drop me a link.

The mode selector. I didn't notice any lubricant on it; I think it would be worth adding some.

Now let's look at the software and the killer feature of these clamp meters.
The software downloads without any trouble directly from the manufacturer's website, and installs just as easily. It takes you straight to the device-pairing menu, where you can filter out all sorts of fitness bands, phones, and other Bluetooth devices.

Once connected to the meter, you can fully control its modes within the range selected with the dial, even turn on the backlight—though it's not clear why you'd need to...
You can also have the readings spoken aloud, in English, and in a rather pleasant voice.
You can also tap the recording mode here and choose the recording interval and number of readings.

The main screen also has extra buttons at the bottom: one displays the current readings as a graph, another shows the data, and then there's... another settings menu, which also has the interval and number of readings. I didn't quite get what this was about at first. You know how it is—you only read the manual after everything's already gone up in smoke, right? :)

So, there are two recording modes. The first is available on the screen without the graph. Tap the button to open the settings menu, choose a mode (or don't), tap Start to record, and... the connection drops. Honestly, I didn't get it at first and was about to start swearing, especially since the app sometimes crashes if you keep tapping around.

If you try to connect to the meter before the set recording interval has elapsed, it offers you the choice of either not connecting or stopping the recording. Aha, now we're getting somewhere.
Anyway, I waited for the recording to finish, opened the app, tapped Record read, and got the recorded data. That is, The meter logs the data without a phone, in standalone mode!
You can view the results either as a zoomable graph or as text.

The menu for viewing measurement results in real time also has a button for recording and saving the results. I also happened to notice that during the recording there were two voltage peaks of around 255 V. I have no idea how the meter registered them or where they came from.
In this mode, data can also be logged at a higher rate, about 3 times per second.
Saved data can be viewed in either offline or online mode.
Standalone mode supports up to 10000 records. At a 1-second interval, that gives you almost 3 hours of recording. Increasing the interval extends the recording time, and the meter's auto-off is disabled while recording.
There is one catch, though: when recording ends, the meter beeps and you can read and save the data. But if you simply open and view the results, they will be lost, so don't forget to save them.
Saving offline recordings isn't entirely intuitive. After recording stops, either automatically or manually, open the meter readings screen, tap Record read, then Save as, and then the floppy disk icon. The data will be downloaded to your phone and available offline. After that, you can view it on your phone or send it by email or through any installed messaging app.

As you can see from the previous screenshot, there's a sharing function. I had no trouble emailing the results to myself in a couple of taps, and they're automatically packed into an archive.
First, an offline recording.

And here's an online recording. You can see that the sampling rate is higher. By the way, I wonder if there's a PC program for working with this data?

A natural question: if the meter can log measurement results, how much power does it consume, and what does that mean for battery life?
Let's take a look.
The top reading is power consumption in voltage measurement mode; the bottom one is in current measurement mode. I didn't check the other modes, since the logger is used much less often there.
In order:
1. The meter's own power consumption: in my view, 1 mA in voltage measurement mode and 7.5 mA in current measurement mode are very economical.
2. Let's turn on Bluetooth—in effect, this is online recording mode. Consumption rises to 7 and 13 mA respectively, so the Bluetooth module draws about 6 mA. Still reasonable.
3. Let's turn on offline recording. Power consumption is less straightforward here, since the current varies during operation: 1.5–7 mA when measuring voltage and 8–17 mA when measuring current.
Now that we've covered power consumption, what does that mean for runtime? Let's do the math.
According to the general specifications, an alkaline AAA battery has a capacity of 1250 mAh, so a set of batteries will last, depending on the mode:
Voltage measurement — 1250 hours
Current measurement — 166 hours
Online voltage recording — 178 hours
Online current recording — 92 hours
Offline voltage recording — over 180 hours; I think the real figure is closer to 250–300
Offline current recording — over 74 hours; I think the real figure is closer to 150
Overall, that's pretty good. But I don't understand why rechargeable batteries aren't used in a device like this. You could charge them either in an external charger or using the meter's own probes, directly from the mains. I don't see why it would be a problem to build in a 1–2 W charger.

Conclusions.
This meter is quite interesting, above all because of its killer feature: the ability to work as a standalone logger. Some time ago, I bought the UT-181A partly for this very reason, and it cost a lot more.
The non-contact current measurement starting at 1 mA is also interesting. Though I suppose you can only really talk about accuracy above 10 mA, the 10 mA–100 A range is still useful for most applications—for example, when I'm working with inverters.
In my opinion, the meter could use MAX/MIN functions and a peak current detector, though the latter is more of a wish list item.
As for accuracy, I was a little disappointed with the voltage measurement mode—I expected better—but it still meets the stated specifications. Maybe I'm getting too picky.
I also recommend applying a little lubricant to the switch contacts; it'll help them last longer.
In short, this is an interesting meter and worth considering, especially the Bluetooth version. I can confidently recommend buying it.
Note: there's a sale on right now, so the basic version costs $32.42 and the Bluetooth version $38.95. At that price, it's worth buying.
That's all for today. I hope you found this useful.
owon cm2100b обзор, https://www.kirich.blog/obzory/izmeritelnoe/1209-tokoizmeritelnye-kleschi-multimetr-owon-cm2100, own 2100b calibration, owon cm2100, cm2100b калибровка, apo в токовых клещах owon, owon сm2100b, cm2100b обзор, owon 2100 разрядность, cm2100b owon, cv2100b, owon cm2100b
Questions and answers
What is the difference between the Owon CM2100 and CM2100B?
The CM2100B has Bluetooth, while the CM2100 does not. Bluetooth enables an app connection and logging during longer measurements.
What current ranges does the Owon CM2100 measure?
It has 2, 20 and 100 A ranges. The author notes that it can measure from 1 mA, but considers readings above 10 mA a more meaningful range for accuracy.
How long can the Owon CM2100B log data independently?
The article gives 92 hours for online current logging and more than 180 hours for offline voltage logging. Offline current logging is specified as more than 74 hours; the author estimates the actual time may be closer to 150 hours.
This model is included in the summary table: Comparison of measuring instruments reviewed on the blog
The figures from this review are in the measurement database: Measuring instruments
Discounted products
You may be interested in
Discounted products
All the content on the site and the advice are free, but we’d be grateful for your support of the project and the channel!
Comments15
Only registered readers can comment. It takes a minute to sign up, and you can log in with the same account you use on the forum.
Log inRegister




Comments are published in their original language.