YR1035, another milliohm meter, an improved version of the YR1030
I’ve already published a review of a similar instrument, where I explained both the theory and practical aspects of the concept of “internal resistance,” as well as the specifics of a four-wire connection. So I’ll try to keep this review relatively short.
We originally planned to order it from Taobao, but the previous shipment had already been packed and sent, and waiting for a new one to be put together would have taken too long. Besides, on 11.11 one of the sellers on AliExpress offered a genuine discount, not a “fake” one.
As it happened, I used this meter for about a month alongside my own, so this review will often compare the two based on hands-on use.
The meter is sold in 7 configurations, with a holder, probes, alligator clips, or combinations of these accessories. I chose the configuration with the meter, probes, and holder.
No complaints about the packaging or delivery: it came in a sturdy cardboard box, with two wrapped packages and a small manual inside.

The contents almost entirely match what I ordered. The only difference is the probes, but more on those later.

The manual is almost entirely in Chinese, which is a shame. The meter isn’t particularly difficult to operate, but a description would still have been useful. The only useful thing in it is a table of internal resistance values for lead-acid batteries.

The probes are noticeably different from mine. They have plastic handles, and the contacts are closer together, which isn’t very convenient.
A set of spare spring-loaded contacts was also included.

The holder, or stand—whichever you prefer—is made of acrylic.

The package includes the most basic version, the BF-1L, made by the same manufacturer as my powerful 100MHz model.

The contact arrangement and connection are fully correct—that is, four-wire, with the contacts connected only through the battery terminal. The holder’s contacts are presumably rated for currents up to 5 A, but even this setup is more than adequate for testing batteries.

The wires are flexible, with silicone insulation, and have a four-pin connector at the end.

Of course, it looks noticeably simpler than my new holder from the same manufacturer, but they’re designed for different jobs. I bought mine specifically for testing batteries under load.

Everything looks very neat, though I had no complaints about the previous one either.
This is where the technical specifications should be, but somehow I couldn’t find them. The most I could find was the information on this page.
Resistance mode resolution: 2200 counts
Voltmeter mode resolution: 100000 counts
The meter measures using AC at a frequency of 1kHz, with a test signal amplitude of up to 100mV rms
Open-circuit probe voltage: up to 2.5Vrms.
Power consumption when off: 5.1uA
Power consumption with the backlight on: 65mA
Power consumption with the backlight on and the probes shorted: 140mA
Power consumption in standby mode: 15mA

Compared with mine, they’re essentially almost identical. The same stiff buttons, the same small screen, and a slightly different front-panel design.
But there are also more substantial changes. The charging port has moved to the bottom edge, and the USB port used to connect the probes has been replaced with a more robust connector. Honestly, though, this is a step backward. Contact resistance doesn’t matter much because it’s a four-wire connection, but the new connector is stiffer and less convenient. All in all, I’m firmly in favor of USB.
There’s a label on the bottom, almost the same as on mine, with a few small differences in the specifications.

My first disappointment—besides the Chinese manual—is that the meter’s menu uses the same Chinese characters. There is an identical device with an English menu, although in my opinion everything is pretty straightforward, and you rarely need to go into the menu anyway. You can easily remember how to turn it off.
Firmware version: ver2.18B2, almost like the temperature sensor marking 18B20 :)

The difference you notice right away after turning it on is that the new version's voltmeter has 2 more digits of resolution than the old one: at startup, we see 5 digits after the decimal point, compared with 3 on mine.

The menu has the same number of items and is not all that different from version 1030. The main difference is that some items are arranged differently.

And here is the menu itself:
Normal mode—I didn't show it in the photos. Press the power button to activate it; press it again to exit. To leave the menu, go to line 1 and press the power button.
1. Component sorting mode.
2. Backlight duration, brightness, and auto-off
3. Time before the device automatically turns off and enters power-saving mode
Auto-off: 5–60 minutes
Power-saving mode: 5–30 minutes
4. Battery sorting mode settings
5. Data transfer speed, apparently for connecting to a computer, although the device has no such port.
6. Ohmmeter and voltmeter calibration. I strongly recommend not touching anything here.
7. Restore factory settings
8. Battery charging current: 200/400 mA; 200 mA by default.
9. Turn off the device from the startup screen by holding down the power button. When the menu shown appears, press the middle button.
It's all in Chinese, but most of the time you set things up once and forget about them. The only exception might be component sorting mode, but it's easier to remember what goes where than to pay extra for the English version of the device.

The device has several modes, with automatic or manual selection of the resistance and voltage measurement ranges. It also has a reading hold mode (hold) and auto-calibration (start it manually by holding down ZERROR).
Again, it's all the same as on my YR1030.

The internals—this is where the differences are more noticeable.

Board version 2.1808, with the manufacturer's marking as well.

The screen can also be removed to make the components easier to access. For obvious reasons, it's best to remove it with the device turned off.

There are clearly more components now. The most noticeable change is the addition of relays, along with more trimmer resistors for adjusting the device.

The board itself now fills the entire case.

Relays. They're probably used to switch the measurement ranges. They sometimes click during operation, but one of them may also click every few seconds while the device is idle (but switched on). It's really annoying in a quiet room.

A 150 V varistor has been added at the device input for protection.

The device is powered by an 18350 cell soldered to the board. Otherwise, it's almost the same as the 1030: the same charger, fuse, and converters, just arranged a little differently on the board.

There's a buzzer on the bottom, and nothing else. By the way, the board is clean, but there’s some kind of strange jumper near the connector.

And here’s a photo of both devices side by side for comparison. I think the difference is more noticeable this way.

The first simple test was measuring the resistance of low-value shunts rated at 1, 2.2, and 10 mΩ. I also tested the shunts themselves in the previous review.

Checking voltage measurement accuracy is a little more complicated. In some ranges, my measuring instruments have lower resolution, and trying to check it with a reference voltage source didn’t work because the meter tries to measure the source’s internal resistance, disrupting its operation.
Still, up to 0.5–0.6 V, the meter has five digits after the decimal point, though most users have no need for that at all.
Above that, its resolution is roughly the same as my Uni-T, although in some places the Uni-T is actually lower.

I also tried checking it at voltages above 30 V, but my power supply has noticeable noise at high voltages, which caused a difference in the readings. It’s small, but it’s there. In this case, it’s more likely a power supply glitch than an issue with the meters.

Another test: measuring the resistance of a shunt connected in series with a capacitor.

First, we measure the capacitor by itself: its ESR is 30 mΩ. Then we connect shunts in series with it and measure the total resistance.
A noticeable difference appeared only with the 1 mΩ shunt: an extra 0.3 mΩ showed up from somewhere. But as far as I’m concerned, that’s insignificant.

Oscilloscope traces of the signals at the meter probes for different resistance ranges. The probe tips are connected together, with the oscilloscope connected to them.

Auto, 20 mΩ, 200 mΩ, 2 Ω, 20 Ω, 200 Ω.
The signal edges are a little more rounded than on the 1030.

Just out of curiosity, I measured the resistance of various batteries. It was only while preparing the photos that I noticed the last four batteries were almost completely discharged—a result of using one of my power banks.
1. Genuine Sony VTC4
2. Fake VTC6
3. Another fake VTC6
4. Xiaomi Ni-MH
5. Soshine LiFe
6, 7, 8, 9. No-name 18650 batteries.

That’s about it for the meter itself, but I’ll say a couple more words about the included holder. It looks simpler than mine, but has one significant drawback: the sides of the case get in the way when inserting a battery—or, more precisely, they make it hard to hold the battery with your fingers. As a result, the battery can simply slip and fall inside the holder.

I eventually got the hang of inserting batteries, but at first it looked more or less like this video. A friend filmed it to demonstrate how awkward the holder is to use.
Since he has to change batteries more often, he asked me to make an adapter so they wouldn’t fall through. I didn’t make it; it was 3D-printed. The adapter fits two battery sizes, 18650 and 14500, depending on how it’s positioned. The photo is also from my friend.

Shown with batteries installed.

Video supplement to the review
Now for a brief summary.
Build quality, calibration, and operation all get top marks. I have nothing bad to say. I’m especially pleased with the measurement range up to 100 V. That’s why I recommended this model to my friend in the first place: he works with high-voltage battery packs more often.
Now for the new features and differences compared with mine:
1. 100 V range versus 28 V, as mentioned above.
2. The voltmeter has five digits after the decimal point, compared with three on mine. In fact, that only applies up to 0.5 V; above that, it’s four digits versus three. But in practice, even three digits are enough in 99% of cases.
3. A microUSB connector on the bottom. In theory, this should be more convenient, but it depends on your preference. Besides, the meter needs charging very rarely, and even if the battery runs low, you can charge it quickly, so it’s not a big deal.
4. The USB connector has been replaced with a sturdier one. In my opinion, this is actually a step backward: the connector is stiff and awkward to use.
5. A standard-size battery. Not great either: if you need to replace it, you’ll have to look for this exact type, and it’s harder to install. A regular flat battery can be replaced with any suitable size.
6. Two relays have been installed inside. Perhaps this improved its operating characteristics, but why do I periodically hear this quiet "click" when the device is just sitting on the table?
7. They added an input varistor, which is useful.
Otherwise, I can say that the YR1030 is sufficient for most applications. But if you work with high-voltage batteries or simply want a more "feature-packed" device, the YR1035 is a good choice. Personally, I'm more than happy with the previous version, too.
yr1035 как узнать емкость от измеренного внутреннего сопротивления, инструкция yr1035, отличия yr1035 rc3563, 2, тестер yr1030 и yr1035 сравнение, yr1035 как поменять щуп, измеритель внутреннего сопротивления yr1035 точность обзор тест, 4108982, чем yr1035 отличается от YAOREA YR1030?, yr2050 yr1035, rc3563 инструкция на русском, yr1035 vs yr1035, настройка yr 1035, 9, kirich миллиомметр, yr1035 обмотка двигателя, kirich yr1035, как пользоваться тестером емкости yr 1035, yr1030 обзор, rc3563 сравнение yr1035, yaorea yr1035 plus инструкция, "обзор" RC3563, yr1035 yr1030 чем отличается, yr1035-4, тестрyr1035, миллиомметр, измерение малых сопротивлений, миллиомметры цифровые, прибор для измерения малого сопротивления
Questions and answers
How does the YR1035 differ from the YR1030?
The YR1035 measures voltage up to 100 V instead of 28 V, and its voltmeter offers higher resolution: up to five decimal places at voltages below 0.5 V. It also has different connectors, a standard-form-factor battery, two relays, and a varistor.
Is the YR1035 suitable for measuring high-voltage battery packs?
Yes. The author recommended it for this use because it measures voltage up to 100 V, while the YR1030 measures up to 28 V.
What test signal does the YR1035 use to measure resistance?
It measures resistance using AC at 1 kHz, with a test-signal level of up to 100 mV rms.
This model is included in the summary table: Comparison of measuring instruments reviewed on the blog
The figures from this review are in the measurement database: Measuring instruments
Discounted products
You may be interested in
Discounted products
All the content on the site and the advice are free, but we’d be grateful for your support of the project and the channel!
Comments14
Only registered readers can comment. It takes a minute to sign up, and you can log in with the same account you use on the forum.
Log inRegister