/       /       /    MUSTOOL X2 Pocket Multimeter Clamp Meter
Support the project on Patreon


MUSTOOL X2 Pocket Multimeter Clamp Meter

$16.99
Go to store
As it happened, I got my hands on a simple clamp meter, which is what I’ll be reviewing today.
I’ll try to keep this brief while still covering the inspection and tests.


This is a fairly budget-friendly instrument, so you shouldn’t expect too much from it. Still, judging by the specs, it can do quite a lot. As it turned out, there’s also a “big brother” model, though it costs quite a bit more. More on that later.

The packaging is very basic, with no color printing, but everything is neatly packed: inside are a case with the meter and a manual.
The picture also seems to show a version without the rotary switch. I wanted to find a link, but couldn’t find one right away, and didn’t see much point in searching too hard.


The package includes:
1. Multimeter
2. Set of test leads
3. Thermocouple
4. Pouch
5. Manual



1, 2. The manual is quite detailed, but unfortunately it’s in English. That said, I think even a beginner can figure out how to use the meter without it.
3. The pouch is just the right size for the meter and has a zipper and a wrist strap.
4. The thermocouple is a standard type. It plugs into the test lead sockets rather than a separate jack, as is usually the case.


The test leads are basic. I didn’t measure the wire gauge, but they don’t feel particularly thick. They’re rated for 1000 V CAT III, but from what I can tell, they wouldn’t actually meet that rating. Still, they’re adequate for household use.


The meter itself looks quite neat. At first glance, I even liked it for some reason.


I didn’t pay attention to its size while handling it, but when I compared it with my main clamp meter, I realized this one is noticeably more compact, despite offering almost the same capabilities.


The top section has a rotary switch for selecting measurement modes and three auxiliary buttons.
This meter comes in two versions. The one I’m reviewing is the X2, but there’s also the X1. The difference is that the latter can measure DC current, but costs noticeably more - link. There are no other differences between them.

As you can see, some switch positions support several modes, which are selected with the Select button.
The Range/Rel button lets you disable autoranging or enable relative measurement mode.
Hold freezes the reading on the display. This button also turns the built-in flashlight on and off.

The switch isn’t too stiff, and it’s easy enough to turn with one finger.



The bottom section has the display and the sockets for the test leads. The sockets are rated for 600 V CAT IV, but I won’t argue about that here. The contacts are recessed quite deeply, and there are only two terminals, so they’ll always be covered by the test leads during use.


The build quality gets a solid four-plus out of five. There’s a small gap between the two halves of the case, and the clamp spring works properly.


The meter is powered by two AAA batteries. The battery compartment cover is secured with a screw, but there’s no retainer to keep the batteries from falling out.
There’s also a flashlight on the back, if you can call a tiny LED that. Its output is roughly comparable to the LEDs used in some lighters. They could have put it on the top edge instead of the back; that would have been much more convenient.

It has an auto-off function. I didn’t time it, but it’s about 10–15 minutes; the meter beeps several times before switching off.


My previous clamp meter opens symmetrically. The design here is a little different, but I can’t say which is more convenient; to me, it makes no difference. A 25–30 mm busbar will fit through the opening without any problem.


The meter’s dimensions. The clamp opening is listed as 32 mm on the right side of the photo, but it’s actually 27 mm. It was simply measured at the widest point in the photo, which is obviously incorrect.


The display is very clear, so there’s little to complain about. Almost all of its features are used in operation. The maximum displayed count is 6000, but in frequency measurement mode it goes up to 9999.
There’s also a backlight, activated by holding down the Select button. It turns off automatically after a few seconds, and its brightness is just right.
When the flashlight is on, a corresponding icon appears in the bottom-right corner of the screen. Unlike the backlight, though, it doesn’t turn off automatically.


Now for the operating modes. The photos below are numbered according to the selector position and operating mode.
1-1. AC voltage, range up to 750 V.
1-2. Frequency when measuring voltage
1-3. Duty cycle
2. DC voltage, up to 1000 V
3-1. Resistance, up to 40 MΩ
3-2. Diode test.
3-3. Audible continuity test, triggers at a resistance below 30–70 Ω
3-4. Capacitance measurement, up to 30000 µF
4-1. Frequency measurement of low-voltage signals, threshold voltage 1 V; peak voltage up to 550 V is claimed.
4-2. Duty cycle of low-voltage signals.
5. Temperature; without a thermocouple, the meter’s internal temperature is displayed
6. Built-in 50–5000 Hz signal generator.
7. AC voltage detector
8. AC current measurement, using the clamp only.
REL. Relative measurements—in essence, a manual zero-setting function.
9. Manual range selection.



Before testing, I decided to take a look inside after all. It comes apart very easily, but I can’t say the same about putting it back together: the clamp mechanism’s spring won’t stay in place on its own, so I had to hold it down while closing the cover.


The inside is very simple. Since there’s no direct-connected current measurement mode, it also feels a bit empty in there.


The “flashlight” LED, buzzer, and clamp connection.


And that’s pretty much everything else. I did try to remove the circuit board from the case, but it turned into one of those “got all the letters right but couldn’t read the word” situations: I removed all the screws, but couldn’t get the board out. It was stubbornly caught on something, and I was afraid to bend it too far because the chip is a blob-on-board type. The only thing I could make out was the display.


I didn’t use any special instruments, calibrators, or anything like that during testing. I measured whatever I had on hand. Next, I’ll quote the specifications for the stated measurement error, followed by the test results.

AC voltage


I tested it over a range of 25–265 V. There’s no point calculating the actual error, since the difference was just one count in the least significant digit, against a tolerance of three counts even before adding the percentage component.


The frequency measurement error isn’t specified when measuring voltage, but I checked it while I was at it.
The reading differed from my multimeter by 0.01 Hz at 50 Hz, but I know my meter reads low by exactly that amount, at least it did when I checked it.


DC voltage measurement.


The first part of the test used a reference voltage source. Its actual output values were 2.499, 5.001, 7.501, and 10.002 V.
The test showed that the results here were also excellent. The only minor issue was at 5 V, but even there the meter passed with plenty of margin, especially since that’s near the end of the range.


For the second part, I simply compared the readings from two meters over a range of 30–61 V. The meter under review consistently read 2–5 counts high in the least significant digits, which is still within the stated error. For example, at 60 V the specified error is 0.5% + 3 counts = 0.33 V; I measured a difference of 0.05 V.


Resistance.


I checked resistance by measuring PTMN-series resistors with 0.25% and 0.5% tolerances. I tested capacitance accuracy in the same way, using both precision capacitors and ordinary ones, whose capacitance I compared with readings from a more accurate multimeter.


Results:
Nominal value – measured value
75 Ω – 75.5
510 Ω – 508.9
1.6 kΩ – 1.597
2 kΩ – 1.995
47 kΩ – 47.1
62 kΩ – 62.0
180 kΩ – 180.3

Assuming the resistors have their exact nominal values, the meter is well within the specified error. Of course, I could recalculate the results using the resistors’ own tolerances, or measure them again with a more accurate meter, but I think that would be overkill.

Capacitance.


Results.
Nominal value – measured value
1000 pF – 1012
1800 pF – 1814
7850 pF – 7952
390.25 nF – 390.6

The error was within the specified limits in every range, even taking the capacitors’ tolerances into account.

Electrolytic capacitors are a different story. Not only are precision capacitors hard to find, but different meters can show different errors when measuring capacitors with the same capacitance. For example, the two capacitors at the top are both rated at 220 µF, but the first reading is off by 9 µF and the second by 7 µF.
Overall, though, it’s good enough for an estimate. For tests like these, I recommend using a dedicated LCR meter, especially since, with switching power supplies now everywhere, a capacitor’s ESR matters more than its capacitance.


Out of curiosity, I checked what was happening across the capacitor terminals during capacitance measurement.
The 1 nF capacitor is on the left, and the 390 nF one is on the right. The frequency is 2 kHz in the first case and about 2 Hz in the second. When testing a 220 µF capacitor, the period was about 1.5 s; I didn’t look any further.


Here’s a clear example of why it’s better to use more specialized instruments to measure capacitors.
The capacitor in the photo is rated at 680 µF. I’ve had it for a long time; it hasn’t been soldered, but I just noticed that it’s bulging slightly.
The meter under review read about 2000 µF. My Uni-T refused to measure it at all, but the LCR meter showed a capacitance of about 60 µF and an ESR of 4 Ω.


Diode test and continuity buzzer


This test checks the voltage across the probes, the polarity, and the current when the probes are shorted together.
1, 4. Semiconductor testing works better with a higher voltage. Here it’s almost 4 V, so there’s apparently a voltage multiplier on the circuit board. The specified current is 1 mA; in reality, it’s about 1.5 mA.
2, 5. Continuity buzzer. The voltage isn’t as critical (it’s 2.1 V here); response time matters more. There’s a slight delay before it beeps, so if you touch the probes together very briefly, you won’t hear anything. I’d call the result average, since I’ve used meters with a longer delay.
3, 6. Resistance measurement: the lower the voltage, the better. This meter applies 0.55 V, and with the probes shorted the current is the same as in continuity mode.


Frequency measurement.


I tested it from 100 Hz to 8 MHz. In every test except the first, the difference was no more than the last digit. At 100 Hz there were two digits, which is perfectly normal given the specified three, even without taking the percentage component into account.
The readings remained stable. By the way, the first and last tests show that the meter can display values above 6000.


Temperature measurement


I checked it against the air temperature and boiling water.
In the first case, both the built-in and external sensors showed the same 26–27 degrees, matching the thermometer. In the second case, the meter showed 98 degrees, which is perfectly reasonable given the stated accuracy. The altitude is about 160 meters above sea level, and the water came from the tap.


A rather unusual feature for a multimeter: a built-in frequency generator. It outputs 50 Hz, then 100 to 1000 Hz in 100 Hz increments, as well as 2, 4, and 5 kHz.
Everything was fine at the lower frequencies. As I mentioned, my meter reads a little low, so the results are consistent.


At higher frequencies, though, things were a little different. At 2 kHz everything was fine, but at 4 and 5 kHz my multimeter started displaying exactly twice the frequency for some reason. And if I connected it through two sets of probes instead of plugging the reviewed meter's probes directly into the Unit's jacks, it even started tripling the reading.


I checked the output with an oscilloscope and found nothing wrong, apart from a little jitter, perhaps. The frequency itself was fine. Most likely, my multimeter's input circuit just didn't like something about it.
Below are the waveforms at 100 Hz, 2, 4, and 5 kHz.


NCV function: non-contact detection of live wires.
In this mode, the display shows either EF, meaning no voltage is detected, or one to four dashes.
1, 2. If you simply place the meter near a wire, it shows 1–2 bars.
3, 4. If you hold it in your hand, sensitivity increases and it shows three bars. Bringing the meter closer gives four bars.
Sensitivity depends on how you hold the meter and whether the probes are connected. Picking it up noticeably increases sensitivity.


AC current measurement


Since I wanted to test it at relatively high currents, I had to put together a makeshift test setup consisting of a Variac, a transformer with a low-voltage secondary winding rated for high current, a multimeter, and, most importantly, a 1 mΩ current shunt with 0.5% accuracy.
This gave me 1 A = 1 mV on the multimeter's display, so there was no need to do any calculations. Only afterward did I realize that, in theory, this setup could measure currents up to 60 A with 1 mA resolution.


I tested it over the 6–60 A range. The specified error is 2% (1.2 A) + 30 counts. My worst result was 0.62 A at 60 A, which is about 1%. Even if the shunt itself is off by 0.5% in the opposite direction, the accuracy still comes out to no worse than 1.5%, versus the specified 2%, even without accounting for the 30 counts.

Unfortunately, tests like these are very inconvenient to run because they require a precision AC current source, which is very expensive :(


I also managed to test part of the upper range, and everything looks fine here too. Unfortunately, I can't go any higher: my shunt is rated for a maximum current of 70 A.


One last test: checking TrueRMS. I connected three multimeters to the output of a UPS: the one under review, one with TrueRMS, and a regular one.
As you can see in the photo, the readings from the meter under review match those from the TrueRMS meter.


Conclusions—and today they'll be mixed, as is often the case.
I have no complaints about measurement accuracy or features—well, almost none. As you may have noticed from the review, the accuracy, even though I checked it using improvised methods, is quite good and often significantly better than specified. The meter is also well equipped; it even has a frequency generator up to 5 kHz, which is a first for me. You might think, just buy it and use it—but there are a few caveats.
Personally, I'm not quite sure who this meter is aimed at. At home, you rarely need to measure current up to 600 A—I'd even say the chances are almost zero. But the need to measure DC current is quite the opposite; for that, though, you need the higher-end version of the meter. The Model X1 costs more, but it sometimes goes on sale for a very reasonable price.
In an industrial setting, this meter could probably come in handy for electricians, for example. But for serious work, you’d be better off replacing the probes. Overall, the meter’s design is more suited to household use, particularly when it comes to safety features.

But speaking subjectively, I liked this little meter, partly because of its size: it’s compact and, surprisingly, very accurate, despite its extremely simple internals. It could really use a DC current measurement function.

That’s all from me. I hope you found this useful.

The store offered a discount for this review: use coupon BGRUMSX2 to get it for $16.99.
$16.99
Go to store
This page was found when searching for:
https://www.kirich.blog/obzory/izmeritelnoe/908-multimetr-tokoizmeritelnye-kleschi-mustool-x2-pocket, mustool клещи, mustool x2 pocket инструкция, китайские токовые клещи плата, световые индикаторы 4 20 ма, световые индикаторы 4-20 ма, мустоол клещи мультиметр отзывы

Questions and answers

Can the MUSTOOL X2 Pocket measure DC current?

No. The X2 measures AC current using its clamp; the higher-end X1 model measures DC current.

How accurate is the MUSTOOL X2 Pocket multimeter?

In the author's tests, readings were close to the reference, and the error was often substantially below the stated tolerance. In the TrueRMS check, the X2 agreed with a meter that has TrueRMS.

What functions does the MUSTOOL X2 Pocket have?

It measures AC current, voltage, resistance, capacitance, frequency and temperature, tests diodes and continuity, and includes a 50–5000 Hz signal generator.


This model is included in the summary table: Comparison of measuring instruments reviewed on the blog

The figures from this review are in the measurement database: Measuring instruments

Calculate manuallyAll calculators

Discounted products


You may be interested in


Discounted products


Support the project

All the content on the site and the advice are free, but we’d be grateful for your support of the project and the channel!

Discuss on the forumHave a question or some experience to share? Discuss the article with the author and readers: the forum is a convenient place to share schematics and photos.
Go to the forum

Only registered readers can comment. It takes a minute to sign up, and you can log in with the same account you use on the forum.

Log inRegister