/       /       /    Advanced XJW01 LCR Meter DIY Kit
Support the project on Patreon


Advanced XJW01 LCR Meter DIY Kit

$ 31.29 (без учета доставки)
Go to store
I've been using a homemade capacitance and ESR meter for quite a while, built from a circuit by GO, a member of the ProRadio forum. I also use another, equally popular meter, the FCL from the cqham website.
Today's review is of a device with higher claimed accuracy that effectively combines both of the meters mentioned above.
Warning: lots of photos, little text. This may be an issue for users on expensive data plans.


I should probably start by mentioning that this device is also sold fully assembled. In this case, though, I deliberately chose the kit: at the very least, it saves a little money, and at best, you simply get the pleasure of putting it together. The latter is probably more important.
I've wanted to replace my previous C-ESR meter for a long time. It basically works, but after at least one repair, it started behaving erratically when measuring ESR. Since I work with switching power supplies a lot (though this matters for conventional ones too), this parameter is even more important to me than capacitance alone.
But this isn't just a C-ESR meter: it measures ESR + LCR, and the full list of measured parameters is even longer. It also claims pretty good accuracy.

Inductance 0.01 uH - 2000H (10 uH)
Capacitance 200pF - 200 mF (10pF) Resolution: 0.01pF
Resistance 2000mΩ- 20MΩ (1.5 Ω) Resolution: 0.1 mΩ
Accuracy 0.3 – 0.5 %
Test signal frequency 100 Hz, 1 kHz, 7.831 kHz
Test voltage 200 mV
Calibration function automatic
Output resistance 40 Ω

The meter can measure -
Q - Quality factor
D - Dissipation factor
Θ - Phase angle
Rp - Equivalent parallel resistance
ESR - Equivalent series resistance
Xp - Equivalent parallel capacitance
Xs - Equivalent series capacitance
Cp - Parallel capacitance
Cs - Series capacitance
Lp - Parallel inductance
Ls - Series inductance

Measurements are made using a bridge method with a four-wire connection to the component.

In my opinion, its closest competitor is the E7-22, but its claimed measurement accuracy is lower (0.5-0.8%), it only has test frequencies of 120 Hz and 1 kHz, and its test voltage is 0.5 V, compared with 0.3%, 120 Hz - 1 kHz - 7.8 kHz, and 0.2 V for the device under review.

This device is sold in several configurations; for this review, I used almost the most complete one. Prices are from the seller's page.
1. Meter only, without enclosure - $21.43
2. Meter + one type of test leads - $25.97
3. Meter + the other type of test leads - $26.75
4. Meter + both types of test leads - $31.29
5. Enclosure for the meter. - $9.70


Everything was packed in a pile of small bags.


Since shipping through a forwarding service usually depends on the parcel's weight, I decided to weigh everything too. Without the cables, it came to 333 grams; with them, quite a bit more: 595 grams.
You can certainly buy it without the cables, especially if you have the parts to make your own. The difference in the kit price alone is about $10, not counting the weight.



I'll start with the cables, by the way.
They're packed in separate bags, and even just by feel, they're quite heavy.


The first set is essentially a pair of ordinary alligator clips, only larger and housed in plastic. But there's more to it than that: the jaws connect to separate wires (connectors) to provide a proper four-wire connection.
The cable is reasonably flexible; most of the stiffness comes from having four shielded wires. The probes connect to the meter using standard BNC connectors, with the shield connected only at the BNC end.

No complaints about the quality. The only thing I didn't like was the lack of color coding by the connectors, since the alligator clips themselves are color-coded. So you have to check which connector goes where every time. The solution: mark them with electrical tape near the connectors.


The second set is much more interesting: it's a tweezer probe for working with small components.
As you can see in the photo, the center conductors are joined some distance from the tweezer tips, rather than right at the ends. So this design is slightly worse than the previous one, but it's also harder to implement the same arrangement as in the alligator clips. There's no color coding.
For easier handling, the tweezers have a guide that keeps the jaws from shifting relative to each other. I don't know how long they'll last, but they're quite convenient to use so far. One thing to note: you need to squeeze them close to the jaws themselves. If you squeeze the tweezers around the middle of the body, the jaws may not close all the way.


A couple of words about what a four-wire connection, or Kelvin connection, actually is. The illustrations are from here, the text is mine :)

When measuring resistance in the usual way (and, by the way, not just resistance), something parasitic like the test leads can have a significant effect. I think many people know that few multimeters show 0 when the probes are shorted and the meter is set to its lowest range. The display usually shows some value around 0.05–0.5 Ω, which is the parasitic resistance.
Sometimes you can compensate for it by enabling the relative measurement function (Rel), but this isn't always convenient, and it certainly isn't always accurate.


The principle behind resistance measurement is quite simple. Connect the component to a current source and measure the voltage across it. But since the wires have resistance, the result will be the sum of the component's actual resistance and the resistance of the wires.
This usually doesn't matter much when measuring high resistance, but when we're talking about values of 1–10 Ω or less, the problem becomes very apparent.
To solve this problem, the current path through the component is separated from the measurement circuit.


In practice, it looks roughly like this, as shown in the diagram.


This method is also used, for example, in power supplies. Here's a photo from my review of a high-power converter. The power circuit and feedback circuit can be separated here too, so the voltage drop in the wires won't affect the voltage at the load.
You've probably also seen something similar in computer power supplies on the 3.3 V rail (orange wires). Only there, a three-wire arrangement is used (that extra thin wire going to the power connector).


A 12 V, 1 A power supply that looks pretty good. I also tried connecting it directly to a load, and it works fine.
But the flat-pin plug makes it inconvenient to use, so I'll replace it with something else. Fortunately, the voltage is standard.
The device can actually run on 9–15 V.
It's a shame you can't choose a package without the power supply. I think many radio hobbyists will have one at home.


The main part of the kit was split into three separate packages.


One of them contained a standard 2004 display (20 characters, 4 lines) with a backlight.


The device's circuit board was carefully wrapped in bubble wrap.


This is one of those cases where the board looks smaller in the store photo than it really is :)
Its actual dimensions are 100 × 138 mm.




The front section of the board is taken up by the connectors for the test probes.


The middle section contains the measurement circuitry, switches, and op-amps. It looks like this section was meant to be shielded, but no shield is included.


The top section contains the “brains” and the power supply.


The first versions of the device used linear voltage regulators; in this version, they've been replaced with switching regulators.
You can also see the power supply connector and the switch.
Replacing the regulators with switching ones can make a noticeable difference when powering the device from batteries. For example, the aluminum case kit includes a holder for three 18650 batteries.


Everything is controlled by a 12C5A60S2 microcontroller. It's based on the rather old 8051 core and has an onboard 8-channel, 10-bit ADC. In the first versions of the device, it used a DIP-40 package; in newer versions, this has been replaced with an SMD part.


There is also a connector on the board for connecting a programmer.


A few close-up photos of the installed components.


The underside is empty; it only has the screen solder pads and test points for the outputs of the voltage regulators and power converters.


And finally, the last bag contains the radio components that still need to be installed on the board.


This includes the keypad board, along with various resistors, capacitors, connectors, and so on.
The design is quite well thought out: the small components are already soldered to the board, so you only need to install and solder the larger ones. In other words, you still get to do some hands-on assembly, but beginners don't have to endure the masochism of soldering tiny components, and it's much harder to mess things up. The result is that you can assemble the device fairly quickly and enjoy the process.


The components are sorted into bags, though most bags contain several different values.


All the resistors included in the kit are precision parts. As a precaution, I measured their actual resistance before getting started.
Assembly is easier because there aren't many different values, and they're easy to measure even with a cheap tester, since none of the resistor values are too close to one another.
The ones at the top are the resistors to solder; there are only six values: 40 Ω, 1, 2, 10, 16, and 100 kΩ.

The resistors at the top are from a labeled bag. They aren't soldered onto the board; they're used to test and calibrate the device. At first I thought they had to be soldered into some critical spots, which is why I measured their resistance. But it turned out they were extras, and the number of resistors to install (16) matches the number in the first bag.


The kit includes capacitors rated at 3.3, 10, 22, and 47 nF, and 0.1, 0.2, and 0.47 µF.
In the photo below, I've labeled the capacitors to match the markings on the board.


A few more parts need to be installed: connectors, a couple of electrolytic capacitors, a relay, and a buzzer.


While waiting for my package, I looked online for more information about the device. It turned out there was not only a schematic, but also different versions of the PCB and firmware, and quite a few people working with this model.
The schematic is, of course, fairly simplified, but it gives you a good general idea.


That also reminded me that about 8–9 years ago, someone in my own city was developing a similar device. Looking at the schematic, you can see a lot of similarities—and it was developed before the one I'm reviewing here.


The seller's comment on the product page really made my day. Sorry about the Google translation.
Put simply (and very loosely), it means: I inspect all the boards and ship them in excellent condition, so don't send me your homemade contraptions, soldered with a hot nail on your kitchen table and phosphoric acid instead of flux.
Love your board and treat it like your favorite girlfriend :)


It's worth noting that both the PCB quality and the component soldering are top-notch. Everything is not only soldered neatly, but has also been thoroughly cleaned!
All the component locations are marked, with both reference designators and component values. Honestly, top marks.


Unboxing video and kit overview.


Let's move on to assembly. When I opened all these bags and laid everything out on the table, I honestly wanted to sit down and solder the whole thing right away. The only thing that stopped me was the decision to put together a short assembly guide in case a beginner wanted to build one.
First, pour the resistors out onto the table and find the ones there are the most of: the 2 kΩ and 10 kΩ values.


Install and solder those first. This will quickly fill most of the empty spaces on the board and make it easier to find the remaining ones later.


I realize this guide is very much for beginners, so I’ll hide the rest of the assembly under a spoiler.


After assembly, we get a rather nice-looking circuit board—as long as we don't mix anything up along the way :)


I shaped the resistor leads with a small jig, but it turned out the lead spacing was a little wider than needed. In the end, I decided to raise the resistors slightly above the board, mostly for looks—I just like it better that way.


After soldering, be sure to clean the board. There wasn't much flux, so alcohol was enough.


After assembling it, I noticed that the board could be shortened a little from its original 138 mm. It can be reduced to about 123–124 mm if you keep the programming connector, or to 114 mm if you cut that out too. In that case, the probe connectors are wired to the holes provided for them. This may come in handy when fitting it into a small enclosure.


The keypad board has only the buttons on it. By mistake, they sent 9 buttons instead of 8—one button had been stuck to another.


On the other hand, they left out one pin header, so I had to rummage around in my stash. I also dug out the matching connectors while I was at it.
In my case, though, I only had right-angle connectors—but plenty of them :)
It's generally useful to keep a set of these connectors around; they come in handy quite often.


Solder the connectors to the keypad board and the display. By the way, the keypad is fully implemented: each button has its own processor pin, rather than using resistors and an ADC, as is sometimes done.


And that's it—the whole kit is ready.


Once assembled, the layout is reminiscent of a multimeter: the display at the top, buttons below it, and connectors below those.


As you can tell from what I wrote above, this is the second, improved version of the instrument. But I like the enclosure from the previous version better, and that's the one I plan to make. The trouble is, that enclosure costs around $9–10, and even more if you buy it with the keypad board and front panel. By the way, I've already reviewed this type of enclosure, when I built an adjustable power supply in one.


My version is designed to be installed in an aluminum enclosure.


And the idea is for it to look like the one in this photo. But design is a matter of personal taste, and I've seen various versions online.


After assembly, I had some test resistors, a button, and a few fasteners left over. And, of course, the power supply and probes.



Now let's move on to the instrument's features and how it works.
When you turn it on, a welcome message appears, followed by the main screen. Everything worked right away, by the way. The instrument has no trimmers at all: assemble it, turn it on, and use it.


The instrument has four main operating modes:
1. Automatic selection. The instrument determines what to measure. It chooses based on the dominant component. That is, if a component is predominantly capacitive, it switches to capacitance measurement; if it's predominantly inductive, it switches to inductance measurement. It can sometimes get it wrong, especially when a component has several pronounced characteristics. For example, some resistors may be identified as inductors.
To complement the automatic mode, manual selection is also available:
2. Capacitance measurement
3. Inductance measurement
4. Resistance measurement.

The display also shows the test signal frequency and measurement range. The ranges are somewhat “unusual,” with no fewer than 16 of them: 1.5, 4.5, 13, 40, 120, 360 Ω; 1, 3, 9, 10, 30, 90, 100, 300, 900 kΩ; and 2.7 MΩ.

By default, the instrument starts in automatic measurement mode at 1 kHz.


A little about the controls.
There are eight labeled buttons below the display.
M - Menu. This is where you perform the necessary calibrations and reset the settings to factory defaults.
RNG - Range. In the menu, this button opens the calibration submenu.
C - Quick automatic calibration.
L - Switches the display mode (first photo). In the menu, it accesses memory.
X - Switches the instrument's operating mode. In the menu, it exits.
R - Decreases the value in calibration mode (X increases it).
Q - Relative measurement mode. Can be used to match two identical components. Connect the reference component, press the button, disconnect the reference, then connect the components you want to match. The screen will show the percentage difference (second photo).
F - Select frequency: 100 Hz - 1 kHz - 7.8 kHz.


The instrument's menu.


Quick calibration by pressing the C button has two options:
1. For capacitance and inductance measurements, the probes are left open.
2. For resistance measurements, the probes are shorted. In both cases, the instrument self-calibrates three times at each frequency.
3, 4. Calibration in resistance mode, showing the probe resistance before and after calibration.


Calibration is quite important when measuring low resistances, since the instrument is sensitive enough to even "see" the resistance of a capacitor's leads, not to mention different wires.



Naturally, this mode is useful for measuring low-value resistors, as well as for "non-standard" measurements such as the contact resistance of buttons, relays, or connectors.


In terms of resistance measurement accuracy, the instrument can easily hold its own against my Unit 181.


The instrument also performed quite well when measuring inductance. The photo shows a 22 µH inductor and three tests of a nominal 150 µH inductor at different frequencies.


Now we can get to the main point—what I mainly need it for: measuring capacitor parameters.

At first, I just tried different capacitors and checked the readings, but one (or rather, a pair) surprised me.
I measured a pair of identical capacitors removed from some old (about 20 years old) Hungarian or Czechoslovak equipment. One measured 488 µF, while the other was almost 600 µF. Nothing unusual there, except that they were originally 470 µF, 40 V capacitors.
What's more, they behave differently at 7.8 kHz. More precisely, the difference in capacitance isn't proportional between them.


Then I took another capacitor (I think it was Matsushita), bought a long time ago but still sitting in my parts stash.
The instrument measured its capacitance correctly at 100 Hz and 1 kHz, but gave a somewhat inaccurate reading at the higher frequency. In general, the instrument sometimes behaves a little strangely at 7.8 kHz, occasionally showing a higher capacitance than at the first two frequencies. Sometimes, when measuring high-capacitance capacitors, it switches to ----OL---- mode or shows an overrange reading above 20 mF.

By the way, the instrument's resolution is good enough to show how the connection point on a lead affects the reading. Even on a single lead, you can see the internal resistance change. The reason I mention this is that people sometimes ask whether a capacitor can be connected using wires if it won't fit in place. You can connect it that way, but its characteristics will be slightly worse.


As you can imagine, just measuring capacitors isn't very interesting, so I borrowed a friend's E7-22. I also noticed that the controls on the two instruments have a lot in common.


First up were film capacitors. At the bottom is a precision 1% capacitor with a stated capacitance of 0.39025 µF.


1, 2. 100 µF polymer capacitor
3, 4. The E7-22 has trouble measuring large capacitances. The instrument under review measures 10000 µF without any problems at 1 kHz, while my E7-22 already showed an overload at 4700 µF.


1, 2. 330 µF CapXon KF series capacitor.
3, 4. A capacitor from the same brand (supposedly), which had simply been sitting in a drawer for several years and has bulged.


This one is just for curiosity's sake. A pair of capacitors from my old motherboard, which ran 24/7 for about 10 years.
1. 2200 µF
2. 1000 µF

The capacitance of the first capacitor has dropped noticeably, but its internal resistance is still fine. More often, it's the other way around: capacitance stays the same while internal resistance increases.


Video of the operation and tests.


If you have any other test suggestions, I have both instruments on hand for now, so I could try them out. The only thing I could think of was checking the test signal's amplitude.
Below is the test signal amplitude relative to ground. The top two traces are from the instrument under review at 100 Hz and 7.8 kHz; the bottom two are from the E7-22 at 120 Hz and 1 kHz. The difference is about 2.5 times.


Above, I mentioned that I planned to use a case where the display would be positioned perpendicular to the surface rather than parallel to it.
But during testing, it turned out that although the display is reasonably good, it is designed to be viewed from the front or from slightly below.


At steep angles, especially when viewed from above or from the side, the image disappears or starts to invert.


That’s why I finally decided to try a display made using VATN technology. I actually wanted an OLED, like the one I used in this review, but a 2004 is almost impossible to buy, and as it turned out, VATN displays are also hard to find online.
In the end, I had to go to a local brick-and-mortar store and buy one there.
There were three models to choose from, with blue, green, and white text. I liked the white one best. The model is WH2004A-SLL-CTV, priced at around $15–16, link. Made by WINSTAR.

At first glance, the displays don’t look very different. At least, the board dimensions are exactly the same: 98 × 60 mm.


There’s a small difference on the underside, but it doesn’t seem significant.


The new display is about 0.5 mm thinner.


The overall connection method is almost the same, apart from a few details I’ll cover below.


First, VATN displays need a negative voltage to adjust the contrast, so the board has a voltage converter based on the well-known 7660, which I’ve also reviewed.
There’s a spot for a trimmer nearby. The center pin goes to the contrast adjustment contact, and the other two go to +5 V and -5 V, respectively.


At first, I wanted to install the trimmer and leave the adjustment entirely to the display board, but then I decided not to remove an extra connector pin. Instead, I simply wired the resistor so that one lead went to the standard contrast adjustment pin (pin 3 on the main connector) and the other to the -5 V output.
I adjusted the image, desoldered the trimmer, and found that I needed a fixed 2.6 kΩ resistor. The closest one I had on hand was 2.49 kΩ, so I soldered that in as a permanent solution.


But that wasn’t all.
Now, pay attention: pin 15 on the connector of a standard display is the positive backlight pin. Here, it’s the negative voltage output, so under no circumstances should you simply swap one display for the other—you’ll just burn it out.

I did things a little differently: I soldered only 14 of the 16 pins.
Pin 16 is the backlight negative, while the positive is connected to the +5 V input, so I simply added a jumper between the backlight negative and the display board’s ground.

And here’s the second warning!
At first, I thought I’d just leave pin 16 in place, since on a standard display it’s connected to the backlight negative. I figured it wouldn’t matter where it was connected to ground. And it would have worked fine, if not for one catch.
The instrument’s display board is powered from +5 V, while the backlight is powered from -5 V. So after connecting the new display this way, I happened to notice just 10–20 seconds later that its backlight was getting extremely hot. I checked it with a multimeter and found that the backlight was getting 10 V, not 5 V (+5 V and -5 V).
So with this instrument, I had to connect the backlight negative to the board’s ground.


Let’s swap the display and give it a try.
Well, it’s not OLED, of course, but it’s a long way from an ordinary LCD, too.
One downside is that it’s designed to be viewed from pretty much any angle except from below. From below, it “blinds” you with a flash.

While I was at it, I measured the current draw with the old and new displays.
1. Old: 48 mA total, or 12 mA for the display alone.
2. New: 153 mA total, or 120 mA for the display alone.

So yes, for a battery-powered device, a standard LCD is much more efficient.


Viewed from above, as I’d planned, it’s easy to read, but inactive pixels start to show up.
It’s easy to get rid of them, but then the display looks dim when viewed straight on, so I settled on a compromise.


The viewing angles are obviously much better than those of a standard LCD; the image is readable even when viewed almost parallel to the screen.
But an interesting effect showed up (in the last photo). If you slowly tilt the screen away from you, at some point (at around 30 degrees) the image fades and starts to invert, then with a little more tilt it almost suddenly looks normal again. So the display is great for vertical installation, but can sometimes be annoying in horizontal orientation.


This is how I intend to use it, and I have no complaints about it in this position.


Next, I planned to give it a permanent home, so I bought a Z1 case. At first glance, it looks neat.


But the case is very large—about one and a half times bigger than necessary—and I was hoping for something more compact.
The case's external dimensions are 188 wide, 70 high, and 197 deep. It's that last dimension I'd like to reduce to 140–150. I could always get out the saw :(
Does anyone know of any suitable cases?



And I guess the review wouldn't be complete without showing what I used until recently.
In the middle is an FCL meter from the cqham website, and on the right is a C-ESR meter by Go from the Pro-radio forum.




There is also a discussion about this instrument, but there's much more information on foreign websites. One Pro-radio user even put together a collection of all the information, firmware, boards, drawings, and so on that he could find. Many thanks to him!

For example, one radio amateur abroad posted a calibration procedure for the instrument.
You won't make sense of it without a few drinks.


That's all for now. I plan to do a short follow-up where I'll finally put all this in a case and share my impressions after using it for a while.

I've been using the instrument for a few days now, and so far my impressions are all positive.
Pros:
1. The satisfaction of building it
2. Excellent PCB and soldering quality.
3. High measurement accuracy
4. It has a 7.8 kHz frequency and a wider measurement range at 1 kHz than the E7-22.
5. Four-wire connection
6. Low power consumption.
7. No need for debugging; the manufacturer claims 0.5% accuracy with basic calibration and 0.3% with manual calibration.
8. A fairly large user community, although mostly overseas.
9. Low price.

Cons:
1. In some situations, the readings at 7.8 kHz aren't quite right. I'll keep testing this.

Overall, I can say that the instrument reviewed here is no worse in functionality or accuracy than the more expensive E7-22, and is probably even better. Of course, there is a difference: the E7-22 can be calibrated and certified, while this instrument is for personal use only.


As usual, I'm looking forward to your questions, advice, test suggestions, and comments. I hope you found the review useful.
$ 31.29 (без учета доставки)
Go to store
This page was found when searching for:
аналоговая lcr приставка, xjw01 корпус своими руками, xjw01 lcr bridge, escort elc-131 инструкция, 1, какие разъёмы для xjw01, конструктор мультиметра diy, руководство по пользователя xjw01, [Обзор] настройка LCR-тестер Xjw01, kit xjw01сборка, kit xjw01 сборка, руководство по экплуатации xjw01, 6, обзор xjw01 lcr meter, тестер xjw01обзор, xjw01 доработки, lcr метр схема, xjw01 - народный измеритель lcr, xjw01 auto lcr, xjw-01, 2, самодельный качественный измеритель lcf, xjw01 прибор для измерения l,c,r, как сделать холдер кельвина для измерение сопротивления аккумуляторов 18650, xjw01 manual, XJW01, LCR-метр, измеритель XJW01, многофункциональный измеритель, мультиметр с LCR метром, набор для сборки


This model is included in the summary table: Comparison of measuring instruments reviewed on the blog

The figures from this review are in the measurement database: Measuring instruments

Calculate manuallyAll calculators

Discounted products


You may be interested in


Discounted products


Support the project

All the content on the site and the advice are free, but we’d be grateful for your support of the project and the channel!

Comments71

Comments are published in their original language.

  1. 18 March 2018 15:23#1
    Читаю ваши обзоры с огромным удовольствием. Спасибо за труд. Уже два устройства купил исключительно прочитав эти статьи.
    +4
    1. 19 March 2018 13:48#2
      Спасибо за поддержку :)
      0
  2. Алексей
    22 March 2018 10:59#3
    Хочу поинтересоваться, какой корпус выбран для измерителя, мобильный или стационарный, по типу RLC2. Также интересно узнать, насколько требователен измеритель к качеству питания 12В. 
    0
    1. 23 March 2018 11:48#4
      Да пока лежит без корпуса, но все таки склоняюсь именно к стационарному, хотя возможно поставлю внутрь аккумуляторы.
      Для питания лучше использовать качественный БП, либо поставить ему по выходу LC фильтр.
      +1
  3. Шататель
    2 April 2018 08:41#5
    "Базовую калибровку", которая компенсация щупов не обязательно в определённых режимах выполнять, можно хоть в авто. Он сам выбирает тип в зависимости от того разомкнуты щупы или замкнуты.
    Причём несмотря на то что при компенсации прогоняются разные частоты, при выполнении её изначально с разных частот - результат немножко разнится. То есть при проведённой компенсации на 1 кГц на остальных значения компенсации будут уже не так актуальны. Если надо что-то на самом дне диапазона ловить - вывод детали, сопротивление мосфета и прочее подобное - влияет ощутимо.

    "Пределы измерения", которых 16 это не пределы, а сопротивление противоположной части моста или что-то такое. Точно не скажу, в принцип работы мостов с автобалансом так глубоко не вникал.
    Пределов там всего 4: 40, 1к, 10к, 100к, по количеству переключаемых резисторов в мосте.
    Пределы можно переключать только в "advanced" режиме, доступном по нажатии кнопки L.

    Как подключать электролиты - всё равно. Прибор без соответствующей приставки вдувать постоянку в них не будет.
    Красное с чёрным просто там просто как обозначение частей с высоким и низким потенциалом.

    Калибровка довольно понятно (для машинного перевода китайский > английский) описана в доках на версию 3.0.
    Но она там более краткая чем в ранних документах, в которых можно почерпнуть некоторые тонкости.
    После калибровки по самопальным эталонам намного больше кондёров стало "попадать в номинал".

    0
    1. 4 April 2018 15:39#6
      Давно пользуетесь прибором? Как впечатление?
      0
      1. Шататель
        6 April 2018 18:32#7
        Пользуюсь с лета 2017 года. Трухольная версия в горизонтальном корпусе с ручкой, плата собрана и калибрована китайцами.

        В целом положительное впечатление, думал хуже будет.

        Примерно с тех пор хочется довести до ума, раздвинуть плату с трансформатором внутри максимально далеко, экранировать корпус, шлейфы экрана и клавиатуры, заменить трансформатор на нормальный, да всё никак руки не доходят.

        Если вы при оформлении в этот чёрный корпус будете ферриты на питание и прочее вешать или ещё что-то - напишите статью, хотя бы кратко что как.

        В коментах тут и на муське никто вроде это чтиво не упоминал, так что сделаю это тут:
        Keysight Technologies Impedance Measurement Handbook - общий букварь
        4263B Service Manual, часть Theory of Operation, если интересно как работает аналоговая часть
        Не помню насчёт 4263B Users Guide, есть там что-то кроме примера того как правильно cцеплять крокодилы при компенсации щупов или нет.
        0
        1. 6 April 2018 19:54#8
          А зачем мучаться с трансформатором, прибор потребляет мало, может лучше заменить его на аккумуляторы и заряжать их раз в год? :)
          Я именно так планирую в итоге сделать.
          0
          1. Шататель
            12 April 2018 20:07#9
            Перепиливать под портатив нет потребности, максимум разъём вывести если потребуется запитать без розетки.
            0
            1. 15 April 2018 20:25#10
              Да, возможно, но иногда нет свободной розетки, потому и думаю насчет портатива.
              0
    2. ШАТАТЕЛЬ
      9 May 2019 23:54#11
      По поводу компенсации, разнящейся на разной частоте.
      Короче если сцепить зажимы встречно как в инструкции - будут ровный нуль омов по всему диапазону частот, но будет не ровной реактивная часть.
      А если сцепить зажимы рядом параллельно на толстый провод - будут вышеописанные артефакты с чуток неровными омами но зато ровная по частотам реактивная часть импеданса. При условии минимального расстояния между проводами и при open и при short компенсации.

      И того и этого сразу достичь не получится и без соединения щупов по Four-terminal pair схеме, что неудобно даже в потенциально модульном варианте.
      +1
  4. Алексей
    5 April 2018 15:11#12
    Проверьте полярность электролита С 311, на фото пустой печатки минус справа, хотя может и не критично.
    +2
    1. 5 April 2018 15:52#13
      Да, действительно, полярность С311 перепутана. Перепаял как надо, спасибо!
      +1
      1. 13 May 2019 18:23#14
        Только увидел, тоже перепаяю :)
        0
  5. kroeger
    11 October 2018 10:20#15
    Здравствуйте! Приобрел LCR метр XJW01 в корпусе. Поработал с ним несколько дней. Решил поиграться с кнопками. Теперь не измеряет малые сопротивления. Боюсь, что и другие измерения будут неправильными. Прошу KIRICH опубликовать соответствие моих кнопок для возврата к заводским настройкам с вашими рекомендациями:

    Нажмите и удерживайте кнопку «M» чтобы попасть в меню (возможно оно работает со второго нажатия).
    Нажмите кнопку «RNG» чтобы попасть в меню калибровок.
    Нажмите кнопку «C» пять раз, чтобы сбросить настройки.
    Нажмите кнопку «L» чтобы сохранить изменения.
    Далее, вернитесь в меню, удерживая кнопку «M».
    Нажмите кнопку «X» чтобы выйти из меню

     
    +1
    1. 12 October 2018 11:09#16
      В целом да, так и есть, но была какая-то мелкая заминка в самом конце, при выходе из меню. Но настройки в итоге все равно сбрасывает.
      0
      1. Kroeger
        12 October 2018 15:16#17
        Просто кнопок "С", "L", "X" у меня нет. Есть "CRL", "S.P/Esr", "QTY/+%". Я вас прошу, напишите порядок сброса до заводских настроек, используя эти кнопки. Думаю, Вас не затруднит.
        0
        1. 12 October 2018 16:09#18
          Плата с кнопками у Вас скорее всего такая же, посмотрите где кнопки на плате из обзора и нажимайте те же у себя не глядя на надписи.
          0
          1. kroeger
            12 October 2018 16:35#19
            Спасибо за помощь. Все получилось. Еле разглядел на Ваших картинках надписи под кнопками. Хорошая у вас подобралась команда. Удачи!

            0
  6. aper13
    13 October 2018 16:10#20
    Здравствуйте, существует ли руководство (на русском языке) по работе с этим прибором?
    0
    1. 18 October 2018 13:37#21
      Мне к сожалению не попадалось, только то что в обзоре и в конце под спойлером.
      0
  7. BOCEM
    19 October 2018 15:54#22
    Приобрёл набор xjw01, собрал, включил, светится цыфиры бегают, а мерить ничего не хочет. Я так понимаю нужно сбросить и калибронуть, вот только не могу найти четкой последовательности нажатья кнопок и действий во время процесса. Судя по видео по началу ваш прибор также себя вёл. Подскажите, пожалуйста, очень хочется почувствовать возможности разработки.
    +1
    1. 20 October 2018 01:59#23
      немного выше в комментариях есть алгоритм сброса настроек.
      0
  8. BOCEM
    20 October 2018 10:55#24
    ДА-ДА согласен надо читать внимательно, на кнопку "М" надо не просто нажать, а и подержатьи тогда всё получается! Спасибо за созданную тему, очень полезна!
    0
  9. Виктор
    25 October 2018 10:10#25
    Здравствуйте. Очень хороший прибор, задумался о его покупке, подскажите пожалуйста он умеет мерять esr без выпаивания с платы? И вообще есть ли какая-то защита от высокого напряжения на проверяемом конденсаторе? Выбираю между ним и esr micro (https://www.radiodevices.ru/esr/esr5.htm) очень хочется услышать Ваш совет... Заранее спасибо...
    0
    1. 29 October 2018 11:06#26
      Защиты насколько я понимаю, здесь нет. по поводу ESR micro не подскажу так как не пользовался им, сравнить не с чем, увы :(
      -1
    2. Александр
      16 January 2019 14:10#27
      из защиты, судя по всему, там только по 2 встречно-параллельно включенных диода 1n4148 на клеммах. лучше все-же разряжать емкости при проверке
      +1
  10. BOCEM
    30 October 2018 08:40#28
    ВИКТОР защита есть выравнивающие диоды на питание, они защищают входа от перегрузки от статики с малых токов убережлт а от сильных зарядов конденсаторов не убережот. 
    0
  11. 4 November 2018 01:09#29
    Очнь интересно, вчера наконец-то спаял :) Все отлично работает.
    А есть ли новая прошивка так и не понял :)
    0
    1. 4 November 2018 02:12#30
      Не, новой прошивки не встречал, да и для обновления нужен как минимум программатор.
      0
      1. Zhuk72
        12 June 2019 17:40#31
        Программатор - обычный USB-UART переходник на CP210x, FT232, CH340 или любом аналогичном чипе. Программа в открытом доступе на сайте производителя.
        +1
  12. 4 November 2018 09:18#32
    Цитата: kirich
    Не, новой прошивки не встречал, да и для обновления нужен как минимум программатор.

    С этими штуками у меня проблем нет, есть целая куча всяких :)
    PS: завелся только с 3-го раза, нужно было очень быстро нажать М для входа в меню и сброса настроек :)
    0
  13. Stepan
    24 November 2018 17:30#33
    По Вашим фотографиям нарисовал схему прибора.По + и - питания стоят по два конденсатора 220 мF x 6,3В.Вероятно по + правый конденсатор впаян не верно(на фото)?
    0
    1. 27 November 2018 17:35#34
      Да, там есть ошибка, я потом исправил, но на самом деле оно заработало и так :)
      0
  14. 30 January 2019 14:13#35
    Здравствуйте.
    Ребята помогите кто может, или ссылкой или советом.
    Купил этот прибор, собрал, включил на экране две строчки квадратиков и всё.
    Ну подумал что контроллер не прошили, начал разбираться в общем обнаружил, что нет сигнала с 3 выхода выхода генератора X201.
    У меня на него приходит почему-то 3.8 вольт.
    Вопрос в следующем каким напряжением питается этот генератор на 32Mhz (у меня на нем написано 32.0-С ASLV0)
    Ну и второй вопрос, где купить.
    Спасибо.


    0
    1. 31 January 2019 17:50#36
      Попробую позже посмотреть что там, хотя конечно брак маловероятен.
      -1
    2. 1 February 2019 04:45#37
      Ребята, всё разобрался.
      Оказался банальный непропай на генераторе (X201) 3 ноги, выпаял, заново запаял всё заработало.

      Теперь другой вопрос, вроде как всё перерыл.

      Где найти информацию по калибровке этого прибора.

      Основную калибровку по резисторам я сделал, а вот как насчет других параметров, что они означают, Z0, PhX, R4b, G2b, и так далее, и самое главное как их откалибровать?

      Спасибо.
      0
      1. KiRiK
        2 February 2019 11:22#38
        посмотри здесь:
        https://drive.google.com/open?id=1Zb3-_jE1ZL-lFmps3UlncTOD-njsyuQa
        в папке LCR Meter XJW01.
        если разберешься, то будет тебе почет и уважение. ибо там много и непонятно.
        0
      2. ШАТАТЕЛЬ
        9 May 2019 23:43#39
        https://www.eevblog.com/forum/testgear/xjw01-lcr-meter-evolution!/?action=dlatt
        ach;attach=326259
        Калибровка и описание параметров.
        Если порыться там на форуме можно найти машинный перевод документа содержащего расширеное описание прибора более ранней версии и его калибровку.
        +1
  15. Sergey
    24 February 2019 22:24#40
    Хороший и доступный наборчик! Я в 1998 за своим ESCORT ELC-131D (готовый прибор,типа E7-22) в Москву ездил - ~300$ !!! тогда стоил.Давно снят с производства и  ему на замуну выпускался более продвинутый АКТАКОМ AM3003(тоже снят с производства) - ~200$! Сейчас это АКТАКОМ  AM-3123 -24000р и AM-3123 -40000р.
    Так что предложенный набор - по цене просто халява!
     Покупал именно из-за возможности измерять Q и D -без них ремонт аппаратуры просто немыслим! Ибо возможность обычного тестера проверить на исправность конденсатора просто никакая!
    0
    1. 25 February 2019 22:06#41
      Да, за свои деньги прибор как по мне отличный, по крайней мере я аналогов пока не знаю.
      0
  16. Stepan
    6 March 2019 13:44#42
    На всех аналогичных схемах, которые удалось найти в интернете,+конденсатора  С124 соединен с анодом диода D109,а на плате данного прибора через резистор R154.Интересно,как эта ошибка сказывается на корректности показаний прибора?
    0
    1. 10 March 2019 04:05#43
      Затрудняюсь сказать, по хорошему надо сравнить оба варианта.
      0
  17. александр
    12 May 2019 20:24#44
    На плате общий вид электролит 220мкф 6.3 вольта правый где стоят в ряд 4 кондера ,полярность перепутана , сравнил со своей платой и методом выпаивания , у меня минус справа, тут слева.
    0
    1. 13 May 2019 14:29#45
      Да, я потом исправил, просто фотоуже не переделывал.
      0
  18. PavelKS
    19 May 2019 19:42#46
    Здравствуйте.
    А Вы обратили внимание, как реализовано двухполярное питание? Из внешнего питания 12VDC с помощью MP1584EN вырабатывается очень точное 10V, а уже из этих 10V с помощью операционного усилителя LM358 и комплиментарных n/p-channel MOSFET на AO4622 вырабатывают +5V, -5V и GND.
    0
    1. PavelKS
      22 May 2019 09:40#47
      Внешний блок питания лучше не менять на "что-нибудь 12V из загашников". Провел эксперимент: подключил к щупам новый конденсатор Rubycon ZLH 1200uF 35V. С блоком питания 12V 2A прибор показал 1245uF 16mOm. Затем переключил на родной 12V 1A без Y-конденсатора, прибор показал 1232uF, 15mOm. Затем переключил на питание от свинцовой АКБ батарейки 12.7V - прибор показал 1223uF, 14mOm.
      0
      1. 4 June 2019 20:13#48
        Кстати именно по этому желательно питать его вообще от трех литиевых аккумуляторов, на странице товара даже был вариант с холдером в комплекте.
        +1
  19. Sali
    7 September 2019 17:19#49
    Приветствую всех!) Собрал данный прибор, от оценок работы пока воздержусь!) Но возник вопрос: можно ли из меню менять яркость и контрастность дисплея или эти параметры зафиксированы аппаратно? В меню данных настроек не нашел!( Подскажите!!)
    0
    1. 8 September 2019 01:32#50
      В меню никаких настроек и нет конечно, регулировка яркости изменением резистора последовательно с подсветкой, а контрастности, изменением номинала делителя по этой цепи. Для контрастности можно поставить подстроечный резистор.
      0
  20. Sali
    8 September 2019 07:31#51
    KIRICH, спасибо за ответ! Я так и думал, но немного смутила Ваша фраза, когда Вы описывали подключение дисплея: "...я хотел установить подстроечный резистор, отдав полностью регулировку плате индикатора, но потом решил не выкусывать лишний контакт разъема и просто включил резистор так, чтобы один контакт шел на стандартный вывод регулировки контрастности" Вот именно слова: "стандартный вывод регулировки контрастности" посеяли сомнения!) Вопрос про регулировки был не праздный, подключил дисплей WH2004L-YYH -CT, все отлично работает, но не хватило штатного напряжения подсветки - оставил питание подсветки от -5v (Катод оставил как было - на -5v, Анод, отключив контакт от платы, подключил через токоограничивающий резистор к земле)  и с контрастностью: напряжение с вывода платы совсем "не попало" - поставил прямо на контакты дисплея подстроечный резистор. А так хотелось обойтись без паяльника и все настройки поделать из Меню..!)
    0
  21. Sali
    8 September 2019 13:14#52
    KIRICH, Вы в ответе писали:
    регулировка яркости изменением резистора последовательно с подсветкой, а контрастности, изменением номинала делителя по этой цепи
    Разобравшись со схемой, выяснил что регулировка контрастности на плате осуществляется не делителем напряжения, а просто резистором подключенным одним концом на землю, а другим на регулировку контрастности дисплея (3-тья нога дисплея). При этом, на экземпляре платы, которая на Ваших фото, номинал резистора R203, который отвечает за уровень контрастности, составляет 2200 Ом, на моем экземпляре платы - 3900 Ом, а номинал R204, который отвечает за уровень подсветки - 1600 Ом. Видимо индикаторы несколько отличаются в разных партиях поставки. В итоге, отключенный от платы 3-ий контакт (регулировку контрастности) дисплея подключил через резистор 1,5 кОм на землю. На мой взгляд получилась отличная контрастность!) Пробовал 1,2 кОм - совсем немного видны незадействованные пикселы, при 1,5 кОм "все чисто"! Пока играюсь с подсветкой, при номинале токоограничивающего резистора 30 Ом свечение дисплея, на мой взгляд, очень слабое, хотя явно заметное по сравнению с резистором в 50 Ом. Ничего меньше 30 Ом под рукой нет, и даже нет 2х50 Ом, так бы подключил в параллель... При 30 Ом в подсветке, ток потребления всего устройства - чуть больше 60 мА.
    0
  22. Sali
    10 September 2019 10:46#53
    Цитата: Sali
    ...номинал R204, который отвечает за уровень подсветки - 1600 Ом.

    Прошу прощения, номинал R204 - 160 Om!
    В итоге пришлось отказаться от использования WH2004L - банально не влез в запланируемый корпус!!)
    0
  23. 7 October 2020 16:01#54
    Здравствуйте
    Купил такой готовый в металлическом корпусе
    https://aliexpress.ru/item/32984209316.html
    Случилась неприятность. Прибор стоял включенный от БП. На корпус случайно попал земляной крокодил от включенного осциллографа. Земля у осциллографа из розетки, может не идеальная но и не от нуля. Прибор выдал нервную морзянку своим реле и перегрузился. После этого стал значительно падать контраст LCD после фиксации показаний, когда реле срабатывает при малых сопротивлениях. В простое и в процессе тестирования детали контраст LCD нормальный. Яркость подсветки при этом все время одинаковая. Показания прочитать можно, но хотелось бы исправить. Смотрел схему, но пока не приходят на ум идеи (
    VDD (с 3.6V до 3.1V) и VSS на экране падают..., но как исправить? А VDD не должно быть 5V?

    Уважаемый Kirich можете подсказать куда копать?

    0
  24. 20 October 2020 18:15#55
    Всем привет. Собрал я данный прибор, он заработал,  но сразу не калибровал. До калибровки он включался, на индикаторе две строчки квадратиков, только со второго включения запускался. Сделал калибровку, стало лучше, но все равно периодически очень редко появляются эти квадратики, и приходиться подождать 30-60 сек потом включить, он запускается. немного поработав, у него почему то стал моргать экран,  как будто плохой контакт, но он в норме, при включений, прога загружена, появляется индикация-но почему то он сам уходит в меню калибровки(думал кнопки-я их от платы отключал-снял плату с кнопками) и тут же перезапускается,появлюется пустой экран даже бес квадратиков, и потомо после 2-3 сек опять перезапускается и отображается это же меню калибровки, питание в норме +5 и -5, не чего не греется. может чего с прошивкой или еще чего, дисплей 2004 проверял отдельно он работает. 
    0
    1. 21 October 2020 10:19#56
      Я бы для начала проверил детали и контакты в цепи регулировки контраста дисплея
      0
      1. 21 October 2020 11:56#57
        резистор добавил на 2,2 к, теперь нормальная яркость, но еще подстрою потом,
        но теперь, стало видно, что он после загрузки программы, уходит сам произвольно в калибровку, и перезапускается и снова уходит в клибровку
        0
  25. 21 October 2020 12:13#58
    Цитата: kirich
    Я бы для начала проверил детали и контакты в цепи регулировки контраста дисплея


    Спасибо за ответ.
    Это Вы на мой вопрос (7 октября, у меня готовый прибор коротнуло землей на металлический корпус и падает контраст на малых сопротивлениях, когда срабатывает реле) или Дмитрию на вчерашний пост (он сам собирал и у него квадратики на экране)?
    Судя по тому что речь про контраст, то думаю мне. Смотрю на схему https://www.kirich.blog/uploads/posts/2018-01/1517162476_38_en_lcr.png - не могу найти где там контраст, в этой старой схеме наверное не было такого. Не могу найти новой схемы для PCB 1.1, CIRC 3.1, 2017 года. В https://www.dropbox.com/sh/grqyruu8u86mzix/AAAhw6FcDBCcJB49OMFdtd54a?dl=0
    не нашел. Не знаете случайно где взять? Или методом тыка придется (((
    0
  26. 21 October 2020 14:46#59
    может с питанием чего то, возможно сделать контраст отдельно мимо схемы, согласно схемам включения LCD

    а мне вот где то прошивку надо доставать, и пытаться прошивать эту штуку

    вам повезло, что еще чип глюки не поймал, после такого, чипы мрут, бывает

    проверил напряжение на кнопках на некоторых 4,9 а на некоторых 0, может и поэтому так происходит либо чегото с чипом
    , если кнопка не нажато то 4,9 если нажата то сигнал 0 на чип
    0
    1. 21 October 2020 15:00#60
      Я бы сначала взял лупу и проверил всю пайку и везде (и тех же кнопок). Проверьте правильность запайки всех электролитических конденсаторов - полярность (и которые уже были с завода). Номиналы и их соответствие у резисторов.
      Перед запайкой деталей на плату, проверяли их исправность ?
      0
  27. 21 October 2020 16:56#61
    Согласен, собирал я свой, все проверял, проблема, постоянного ухода в меню калибровки, была в разъёме, куда вставлял плату с кнопками, при пайке, туда как то попала канифоль, хотя я очень мало олова наносил, убрал этот разъём почистил. Больше не уходит в меню сам по себе, остальное работает. 
    0
    1. 21 October 2020 18:27#62
      НУ вот, как и ожидалось - проблема в сборке.
      Я как-то давненько тоже паял ДИУ - и (как оказалось) перегрел один из переключателей. Оно вроде и работало, но как-то было все не так, как надо. Тоже искал, в чем траблы...
      В таких наборах обычно все ок, если сам не накосячишь.

      Кстати, в этом  XJW01 в первые дни у меня были непонятки с измерением емкости на разных частотах, да и просто быстрая калибровка по резисторам ощутимо плавала около нуля. Сделал полную калибровку, и начал обращать внимание, на какую поверхность ставлю прибор - и вопросы ушли, все стало стабильно и здраво.
      0
  28. 21 October 2020 18:35#63
    Сори за не в тему :)
    Кирич, я Вам написал в личку на майску по поводу БП для лабораторника, если не трудно, ответьте пожалуйста :)
    0
    1. 22 October 2020 12:04#64
      Приветствую.
      Очень много пишут, если не сложно, напомните пожалуйста о чем речь, не могу найти Ваше письмо.
      0
      1. 22 October 2020 13:57#65
        На майску отписался :)
        0
  29. 22 October 2020 12:11#66
    имя владельца письма, которого вы хотите найти?
    0
    1. 22 October 2020 12:27#67
      Это был вопрос к KIR, у меня на сайте уведомления приходят всем кто участвовал в ветке комментариев. Надо будет как-то перенастроить, а то не всегда удобно и сбивает с толку.
      0
  30. 22 October 2020 12:52#68
    интересно о обмотки трехфазного асинхронного электродвигателя на 220/380В разной мощности можно мерить этим прибором  LCR-метра XJW01-естественно что питание отключаем, тк там индуктивное сопротивление, но оно очень низкое где то от 1-10 ом одной обмотки, это на случай определение межвиткового замыкания, или на земля кз определить, но обучно мегометром пользуются конечно.
    0
    1. 21 December 2020 03:22#69
      Хороший вопрос, странно что не заметил раньше. Надо будет попробовать просто на каком нибудь двигателе или 50Гц трансформаторе.
      0
  31. 22 October 2020 16:57#70
    Нашел схему Schematic_STC12_LCR_2004LCD_20200302173656.pdf на cristalradio:
    http://www.crystalradio.cn/forum.php?mod=attachment&aid=MTI4Njg2NzZ8MjUyYWM
    2ODV8MTYwMzM3NDQyNXwwfDIzMTkzMw%3D%3D
    Может кому пригодится.
    0
    1. 21 December 2020 03:24#71
      спасибо, для более удобного добавления ссылок можно делать их при помощи "цепочки", тогда ссылка будет в нужном слове.
      0

Discuss on the forumHave a question or some experience to share? Discuss the article with the author and readers: the forum is a convenient place to share schematics and photos.
Go to the forum

Only registered readers can comment. It takes a minute to sign up, and you can log in with the same account you use on the forum.

Log inRegister